3S-CL Zn Zinc Ion Online Analyzer is an analytical instrument used for online analysis and process monitoring. It uses LED light source for colorimetric analysis and measurementZinc ionConcentration.
Zinc reacts with reagent zinc in a buffered alkaline solution to form a blue complex.The absorption intensity is proportional to the zinc concentration in the sample and measured at 660nm.
Product advantages
The instrument adopts a dual compartment housing to ensure complete separation between the electronic components (upper part) and the sampling and analysis components (lower part).
The instrument uses less reagent consumption, reducing operating costs.
The instrument can use an internal dilution module, with a measurement range ranging from trace μ g/L to 1000mg/L; Multiple chemicals can be supplied to meet actual needs.
The analyzer is equipped with a temperature heating module.
The analyzer can be equipped with a refrigerator to keep the chemicals at the appropriate temperature.
The equipment can be set up for automatic calibration, verification, and zeroing, which can reduce downtime and operator intervention, ensuring accurate results.
The analyzer adopts a color touch screen interface to display measurement values and status information. Easy to access menus and functional interfaces. And supports USB data download.
You can choose the dual channel version. By adding an external sequencer, the number of sampling channels can be increased to 4.
The instrument undergoes factory testing before leaving the factory. Simply connect the power supply, sample and reagent pipelines, and the analyzer can operate.
Product Parameters
measurement parameters |
Zn2+(ppb, ppm, mg/L) |
size |
604*380*242mm(H*W*D) |
Measuring principle |
Zinc reagent spectrophotometric method |
weight |
44 pounds (20 kilograms) |
measurement range |
0.01-1ppm Zn2+For a 26mm reaction tank 0.02-2.5ppm Zn2+For 16mm reaction tanks; Up to 125ppm of Zn2+with optional internal dilution |
power supply |
110/220V AC 50/60Hz or 24 VDC Power consumption: MAX80 watts |
通道 |
1,2 adopts built-in switch valve 3,4 use external sequencers |
output |
Two 4-20mA analog signals; One Modbus RTU RS485 digital signal; |
frequency analysis |
Freely programmable, close to continuous analysis |
call the police |
2 programmable SPDT passive relays |
cycle time |
6-8 minutes, including preparation before the analysis cycle and rinsing time after measurement |
digital input |
Remote start/stop operation |
reactor |
Built in temperature heating module |
temperature |
41-113℉ (5-45℃) |
water sample |
Temperature: 5-50 ° C Connection interface: 6mm Flow rate: 80-500mL/min |
humidity |
10% to 90% non condensing (indoor installation); Outdoor installation requires the use of protective cabinets or rainproof systems; |
drainage |
No pressure, standard atmospheric pressure drainage Connection: 12mm (1/2 inch) |
install |
Wall mounted (standard), Desktop support or panel installation (Option) |