ZDCLY-6E microcomputer fast integrated sulfur

1Adopting standards:
GB/T 214-2007 "Method for determination of total sulfur in coal"
GB/T 483-2007 General Provisions for Coal Analysis Experimental Methods
GB/T 387-1990 "Determination of sulfur content in dark petroleum products - Tube furnace method"
II Scope of application:
Suitable for the determination of total sulfur in materials such as lignite, bituminous coal, anthracite, coke, dry coal samples of coal water slurry, and heavy oil in industries such as electricity, coal mining, metallurgy, chemical engineering, cement, geological exploration, and scientific research institutions.
III Technical parameters:
1. Sulfur measurement range: 0-100%
2. Heating time: ≤ 25 minutes
3. Analysis time: about 3-5 minutes
4. Resolution: 0.001%
4performancecharacteristic:
1. Use a touch sensitive film panel to input data, and control functions with a touch keyboard. The operation is simple and convenient.
2. Set furnace temperature and repairThe sulfur content can be entered through the keyboard without opening the panel.
3. Automatically determining the endpoint of titration has shortened the experimental time.
4. Adopting PID control of furnace flow and automatic adjustment of furnace flow size can extend the service life of silicon carbon tubes compared to traditional controllers.
(Automatic control of current from small to large)
5. The sample delivery mechanism adopts electronic switches, which have higher reliability.
6. There will be no over electrolysis phenomenon due to the sequence of on/off.
7. Automatically turn on electrolysis.
8. The integrated design has a high degree of automation and a small size, making it an ideal sulfur analyzer for laboratories at present.