16The input channel is optional.
■ 0.2Grade accuracy, measurement and control speed per channel0.1In seconds.
■ All channels are universal inputs that can be selected through parameters such as thermocouples, thermistors, DC current, or DC voltage signals.
■ standard configuration2Point public relay output, for all input channels alarm output or break resistance, break couple alarm; It can also be selected as an independent alarm output for a certain channel.
■ standard configuration1roadRS485The communication interface can be selected as a transmission output, but transmission and communication cannot be selected simultaneously.
content |
Code Description |
||||||||
XSLC- |
0.2Level measurement accuracy; Measurement and control speed(0.1X number of channels) |
||||||||
input channel |
08 |
8Road input |
|||||||
16 |
16Road input |
||||||||
Overall dimensions |
160×80×124 horizontal format Hole size:152×76 |
||||||||
Panel specifications |
2bitLED(Channel number display)+ 4bitLED(Measurement value display) |
||||||||
input signal |
All channels are universal inputs, and the input signal is set by parameters: Thermal resistance(Pt100/Cu100/Cu50/BA1/BA2/G53) Thermocouple(K/S/R/B/N/E/J/T) Direct current((4-20)mA/(0-10)mA/(0-20)mA) DC voltage((0-5)V、 (1-5)V)Signal or(mVSignal(+100 ~ -100)mV) |
||||||||
call the police |
2Click on the public alarm. (Relay output,250V AC/3A, resistive load) |
||||||||
Communication (communication speed:2400、4800、9600、19200、38400、57600)(Mailing address:0~99) |
S1 |
RS232Interface, selected through parameter settingsTC ASCII orModbus-RTUagreement |
|||||||
S2 |
RS485Interface, selected through parameter settingsTC ASCII orModbus-RTUagreement |
||||||||
|
Transmission (optional) (cannot be provided simultaneously with communication. Resolution:1/3000Load capacity:600Ω)
|
A1 |
Current output(4-20)mA、 (0-20)mA、 (0-10)mA |
|||||||
A2 |
Voltage output(0-5)V、 (1-5)V |
||||||||
Power Specifications |
V0 |
100-240V AC 50/60 Hz |
|||||||