Usage Overview: Used for online measurement of turbidity in water bodies PH、 Residual chlorine content. It can be widely used for on-site online monitoring in water plants, power plants, industrial water use and other departments.
1. Microcomputer, Chinese operating interface, large screen LCD with backlight display screen, can simultaneously display clock, parameters, data, units, status prompts, etc;
2. Integrated box design, integrated stable flow defoaming turbidity measurement sensor, integrated flow channel, equipped with inlet pressure and water flow regulation;
3. Equipped with a long-life light source, a low-power chip, a low drift, high-precision circuit system, power down saving function, and a data output interface;
4. Reliable positioning structure and high-precision optical path system, fast response speed, accurate measurement, effective chromaticity compensation, and direct reading of turbidity values;
5. Equipped with user panel calibration and automatic factory calibration status restoration functions, it has simple operation and can effectively ensure the accuracy of measurement data;
6. Wall mounted installation, easy maintenance, and the instrument can work stably for a long time.
PRODUCT MODEL |
WGZ-1002C Water quality analyzer (single parameter) |
WGZ-503C Water Quality Analyzer (Three Parameters) |
|||
Measurement parameters |
turbidity |
PH |
turbidity |
PH |
Residual chlorine |
measuring method |
860nm LED light source 90 ° surface scattered light turbidity |
electrode |
860nm LED light source 90 ° surface scattered light turbidity |
electrode |
Membrane electrode (imported) |
Water sample import method |
Water collection method |
Constant current flow channel |
Water collection method |
Constant current flow channel |
Constant current flow channel |
measuring range |
1~1000NTU |
0~14PH |
0~20NTU |
0~14PH |
0~2mg/l |
Minimum display value |
1NTU |
0.01PH |
0.001NTU |
0.01PH |
0.01mg/l |
Water sample working flow rate |
500~1500ml/min |
500-700ml/min (equipped with flow meter to adjust flow rate) |
|||
Water inlet pressure |
0.05-0.6 MPa (equipped with inlet pressure valve regulation) |
||||
working conditions |
The water sample temperature is between 4-40 ℃, the ambient temperature is between 0-45 ℃, and the ambient humidity is not greater than 90 ℃% RH without condensation |
||||
output signal |
Optional communication methods such as 4-20mA, RS485, etc. (should be specified when ordering) |
||||
power supply |
DC24V/3A or AC220V/50Hz/20W (to be specified when ordering) |
||||
Complete configuration |
Sensors and displays |
Online continuous measurement and display of three parameters simultaneously, equipped with inlet pressure valve, flow meter adjustment, flow channel, automatic defoaming design, long-life light source, automatic range switching, RS485 signal output. |
Note: The measurement range and parameters can be customized according to user requirements, including single channel or dual channel wall mounted or embedded instruments, flow type, pipeline type, and input type sensors.