Ultra low range online turbidity analyzer Monitor AMI Turbitrace
Application field: High precision turbidity meter used for measuring turbidity in pure water and sewage
Product details

Application fields:

High precision turbidity meter is used to measure turbidity in pure water and wastewater. The water quality monitoring of tap water before filtration, after filtration, sedimentation effluent and factory water, as well as municipal pipeline network, complies with ISO7027 (EN27027/DIN38404)

Instrument features:

· Zero point automatic programming

· A complete half micron molecular filter can achieve zero drift compensation measurement

· The instrument includes a transmitter, turbidity electrode, flow electrode, and a flow cell with a half micron filter. All components are fixed on one panel

· Valves and connectors used for slope calibration comply with ISO 7027 standard

· Large screen backlight displays measurement values, flow rates, and operational status

· Full text menu driven user serial port mode

· Two channels of 0/4-20 mA analog output, isolated from electrode input current

· Integrated alarm dry node, programmable alarm values and instrument faults

· The alarm functions include: "Flow rate exceeding limit", "Request cleaning", and "Filter replacement"

· Programmable limit switch or PID control for two dry junction points

· Dry node input, programmable function

Technical Specifications:

· Turbidity meter with flow electrode and needle valve

· Range: 0.000-100.0 FNU

· Accuracy: ± 0.001 FNU or 1% of the reading

· Response time: T when the sample water enters at 10 l/h90 <15 seconds

· Operating temperature: 1-40oC

· Sample water flow rate: 5-20 l/h

· Sample water pressure:1-10 bar

· Sample water inlet:Scepter 6 mm

· Sample water outlet: no pressure discharge (funnel connected with 15x20 mm hose)

· Transmitter: Aluminum IP 66

· Transmitter size:180x140x70mm

· installer versionSize: 850 x 280 x 200 mm

· Installation disk: PVC

· Weight: 8.5 kg