Shenzhen Sansheng Zongheng Technology Co., Ltd
Home>Products>Think twice about VTM1000 series hot deformation Vicat softening point testing machine
Product Groups
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
  • Address
    Block C, Building 1, No. 38, Youmagang Industrial Zone Avenue, Shiwei Community, Matian Street, Guangming District, Shenzhen City
Contact Now
Think twice about VTM1000 series hot deformation Vicat softening point testing machine
Think twice about VTM1000 series hot deformation Vicat softening point testing machine
Product details

Product overview

Think twice about VTM1000 series hot deformation Vicat softening point testing machine

Product Usage:

Purpose: Used to measure the thermal deformation and Vicat softening point temperature of thermoplastic materials and thermoplastic pipe fittings.

Standards
GB/T 8802-2001 "Determination of Vicat softening temperature of thermoplastic pipes and fittings"
GB/T 1633-2000 "Determination of Vicat Softening Temperature (VST) of Thermoplastic Plastics"
GB/T 1634.1-2004 "Determination of plastic deformation temperature under load - Part 1: General test method"
GB/T 1634.2-2004 "Determination of Load Deformation Temperature of Plastics Part 2: Plastics, Hard Rubber and Long Fiber Reinforced Composite Materials"
GB/T 1634.3-2004 "Determination of plastic deformation temperature under load - Part 3: High strength thermosetting laminates"

Performance characteristics

Using a high-precision digital micrometer imported from Switzerland as the deformation measuring element for the sample, with high accuracy and good stability, fully ensuring the accuracy of the test data;
The connecting rod of the sample holder and the load rod are made of low expansion alloy with the same diameter and material, which can minimize the deformation of the sample holder and improve the accuracy of the test results;
◎ Equipped with automatic lifting function for the sample rack, driven by a linear motor with strong load-bearing capacity and smooth operation;
Each sample is equipped with a temperature sensor PT100 to ensure the accuracy of the experimental results;
◎ Can simultaneously perform two types of tests: thermal deformation and Vicat softening point temperature;
◎ There are water cooling, natural cooling and other methods to choose from according to user needs;
Equipped with multiple safety protection devices to ensure the safety of testing equipment and operators;
Specialized testing software with powerful functions, user-friendly interface, and easy operation.

Technical Parameter

model VTM1300 VTM1400 VTM1600
Temperature control range Room temperature -300 ℃
Uniform heating rate A speed: 5 ± 0.5 ℃/6 min B speed: 12 ± 1.0 ℃/6 min
Using media Silicone oil or transformer oil (user provided) 30L
Deformation measurement range >10mm
Deformation resolution 0.001mm
Number of sample racks 3 4 6
Control temperature sensor PT100 3 4 6
load (1) Vika load: GA=10N ± 0.2N GB=50N ± 1N;

(2) Thermal deformation bending normal stress:

Method A, using fiber stress of 1.80Mpa;

Method B, using fiber stress of 0.45Mpa;

Method C, using fiber stress of 8.00Mpa

Maximum heating power ≤5KW
power supply AC three-phase five wire 380V ± 10% 50Hz
Cooling method Natural cooling and water cooling
Special Features Equipped with upper limit temperature protection function setting

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!