1TW-6566 Sodium Ion Monitor (Steam)brief introduction
Continuous monitoring of trace sodium ions in ultrapure water can be widely used in industries such as power, petroleum, and chemical. It mainly monitors the content of sodium ions in the main steam pipes and saturated steam of power plants, and has the characteristics of fast reaction, accurate measurement, and easy operation.
IITW-6566 Sodium Ion Monitor (Steam)main features
Advanced embedded microcontroller technology;
High quality imported electrodes, effective alkalization method, and automatic stability judgment function ensure the stability and reliability of measurement
Advanced surface mount technology and integrated design improve cost-effectiveness and reliability
Exquisite structure, disc installation, all aluminum frame box, beautiful and sturdy, strong anti-interference ability
Simultaneous display of multiple parameters, rich content, easy to understand
Ø Automatic or manual temperature compensation
Ø Two way isolated current output and two way alarm signal
Ø Adopting standard RS485 digital communication
Ø With large capacity historical data storage and historical curve query function
3、 Main technical performance
1. Electronic unit
Ø Display: 192 × 64 dot matrix LCD, can choose to display Chinese or English
Ø Measurement range: (0.00~9.36)pNa,0.01ug/L~23.0g/L,(-1000~+1000)mV
Ø Resolution: 0.01pNa,0.01μg/L,0.1mV
Basic error: ±0.02pNa
Stability: ±0.02pNa/24h
Ø Current output: (4-20) mA (dual isolated output)
Output accuracy: ±1%F.S
Output load resistance: 800 Ω
Ø Power supply: AC(85~265)V, Frequency (45-65) Hz
Ø Power consumption: ≤ 10W
Ø Insulation strength: Power supply to ground 1500V/1 minute without arcing
Ø Alarm output: Two channel alarm output, programmable alarm content
Ø Communication: RS485 (optional)
Ø Automatic data storage: can be set to automatically record historical data, historical events, and calibration records
2. Whole machine
Ø Measurement range: (1.00~8.00)pNa, 0.23ug/L~2300mg/L
Ø Resolution: 0.01pNa or 0.01 μ g/L
Basic error: ± 0.05pNa or ± 5 μ g/L, take the larger value (within ± 10 ℃ of the calibration temperature)
Response time (T90): 2min (25 ℃)
Ø Temperature sensor: Pt1000
Ø Temperature measurement range: (0.0~99.9)℃
Ø Temperature measurement accuracy: ±0.5℃
Ø Temperature resolution: 0.1℃
Ø Temperature compensation range: (0.1-60.0) ℃ (manual or automatic)
Ø Calibration methods: sample calibration, standard solution calibration
Ø Dimensions: 450mm x 690mm x 190mm (width x height x depth)
Ø Hole size: 410mm×645mm
Ø Weight: 20kg
3. Sample requirements
Ø Temperature: (5-50) ℃
Ø Pressure: (50~300)kPa
Ø Traffic: (6~20)L/h
Ø The allowed solid content of the water sample is ≤ 5 microns (no gelatinous substances or oil are allowed)
4. Sample requirements
Environmental temperature: (5-45) ℃
Environmental humidity: ≤ 90% RH (non condensing)
Storage and transportation temperature: (-20~55) ℃ (excluding electrodes, electrodes should be above 0 ℃)
4、 Configuration
Ø One online sodium measuring electrode
Ø One online sodium reference electrode
One bottle of sodium standard solution
Ø One bottle of reference electrode filling solution
Ø One bottle of electrode activation solution
Ø Instrument door key