TC830 Coating Thickness Gauge
TC830 Coating Thickness Gauge
Product details

TC830Portable intelligent digital coating thickness gauge, which can quickly, non destructively, and accurately measure the thickness of coatings and coatings. It can be used in both laboratories and engineering sites. This instrument can be widely used in manufacturing, metal processing industry, chemical industry, commercial inspection and other testing fields. It is an essential instrument for material protection professionals.

Execution standards

1. GB/T 49561985 Measurement of thickness of non-magnetic coating on magnetic metal substrate using magnetic methods

2. GB/T 49571985 Measurement of non-conductive coating thickness on non-magnetic metal substrate using eddy current method

3. JB/T 83931996 Magnetic and eddy current coating thickness measuring instrument

4. JJG 88995 Magnetic Resistance Thickness Gauge

5. JJG 81893 Eddy current thickness gauge

TC830Type coating thickness gauge is a substituteCMSeriesMCSeriesTTSeriesFTSeriesCTSeriesHCCSeries

Surfix GermanyThe most ideal cost-effective products such as series

Instrument features:
1.
Using magnetic sensors to measure the thickness of non-magnetic coatings on magnetic metal substrates

Using eddy current sensors to measure the thickness of non-conductive coatings on non-magnetic metal substrates
2.
Has two measurement modes: high-precision single measurement mode(Precision)And scanning mode(Scan

3. Equipped with temperature compensation function: Advanced real-time temperature compensation technology can automatically detect changes in environmental temperature and probe temperature caused by them

Compensate for measurement errors to make measurements more accurate

4. FAfter the wear and tear of the magnetic sensor, it can be automatically matched with the instrument host again. After successful matching and basic calibration, it still remains

Can continue to be used, thereby extending the lifespan of the sensor
5.
There are five statistical measures: mean(MEAN), maximum value(MAX)Minimum value(MIN)Test frequency(NO.)The

Standard deviation(S.DEV
6.
The instrument can be calibrated using zero point calibration, one point calibration, or two-point calibration methods, and basic calibration methods and temperature systems can be used

Correction of systematic errors in sensors using numerical calibration method
7.
Has storage function: can store up to500A measurement value
8.
It has a deletion function: it can delete individual suspicious data that appears during measurement, as well as all data in the storage area,

In order to conduct new measurements
9.
Can set limit: automatic alarm for measurement values outside the limit
10.
Equipped with power and battery level display function
11.
During the operation, there will be a beep sound prompt and the buzzer can be turned off

12.There are three shutdown modes available: manual shutdown mode, timeout automatic shutdown mode, and low battery automatic shutdown mode

Set timeout automatic shutdown waiting time, accompanied by backlight flashing and buzzing sound prompts during automatic shutdown

13.Real time data communication function: can send real-time measurement results toPCEditing or printing on board

Technical parameters:

model

TC830FNmagnetism/Eddy current dual-use type

Working principle

TC830Fmagnetic induction

TC830Neddy

measuring range

01500um(Other specifications can be customized)

Resolution

0.1um (Measurement range less than100umtime)

1um (Measurement range greater than100umtime)

Calibration of indication point

±(2%H+1)

Two point calibration of indication value

±[(12)%H+1]

Basic calibration

±(1%H+1

Minimum measurement area

Φ7mm

Φ5mm

Thinnest measuring substrate

0.5mm

0.3mm

Minimum curvature radius of the substrate

Convex:1.5 Concave:10mm

Convex:3 Concave:10mm

the metric system/English system

You can choose between metric and imperial units

Battery voltage indication

There is a low voltage prompt function

Automatic shutdown

Has automatic shutdown function

Having data communication capability

Can send real-time measurement results toPCPrinting on board

electricity source

three links1.5VNo.7 dry battery

Usage environment

Temperature:040 Humidity:1090%RH

External dimensions

110×50×23mm

weight measure

100g(Excluding battery)

TC830FNDual use standard configuration:

name

quantity

name

quantity

host

1tower

Standard test module

5a flat

Magnetic Sensor

1only

7Dry battery

three links

eddy-current sensor

1only

operation instructions

1stem or root of plants

Iron based reference block

1piece

Certificate of conformity, warranty card

1share

Aluminum based reference block

1piece

Instrument packaging box

1only

■ After sales service:3Free exchange within one month, free two-year warranty, lifelong maintenance