Shimadzu LC-10A Dual Pump Liquid Chromatography System
LC-10ATVPPlus infusion unit
Corresponding to conventional analysis, the infusion unit improves maintenance simplicity, analysis reliability, and instrument durability while achieving high performance, reducing maintenance frequency
Adopting a new type of plunger, reinforced PTFE sealing ring, and plunger floating installation method that considers material structure, greatly improves the durability of consumables.
·Flexibility
By adding necessary units to the basic infusion unit, it can be expanded into a high-pressure gradient. The gradient control can also be performed by the infusion unit host.
·Validation support
Record maintenance information and operation records on the host, which can be read from the VP key. And it has safety protection during calibration, supporting GLP/GMP
Shimadzu LC-10A Dual Pump Liquid Chromatography System
LC-10ATVPPlus
Infusion method: Micro volume series double plunger
The maximum infusion pressure is 40Mpa
What is the flow setting range of 0.001? 9.999mL/min
Flow accuracy ≤ 0.1% RSD
Gradient concentration accuracy ± 1% (0-100%, 2-liquid gradient of water/acetone aqueous solution)
Dimensions W260 × H140 × D420mm
Weight 11Kg
Operating temperature range: 4-35 ℃
Shimadzu LC-10A Dual Pump Liquid Chromatography System
UV Visible Dual Wavelength Detector SPD-10AVPPlus
A high-sensitivity detection unit with excellent signal-to-noise ratio and rich functions such as dual wavelength simultaneous measurement.
·Extremely high low noise level and baseline stability improve sensitivity
By using efficient optical systems, digital filters, etc., a noise level of ± 0.35 × 10-5AU has been achieved. In addition, the lamp chamber and monochromator are completely separated, and the lamp chamber has cooling functions to prevent the temperature rise of the monochromator and improve baseline stability.
·Dual wavelength simultaneous measurement and proportional chromatography function
The dual wavelength simultaneous measurement function improves the efficiency of multi-component system analysis. By using the proportional chromatogram of the ratio of dual wavelength absorbance, impurities can be detected and peak purity information can be obtained.
·Time program and wavelength scanning function
When switching wavelengths using a time program, eliminate baseline drift through the baseline adjustment function. In addition, the wavelength scanned spectrum can be saved in the workstation and compared with the standard substance spectrum for confirmation.
·Equipped with long-life deuterium lamps
By utilizing the time program function, the deuterium lamp can be automatically turned on/off to extend its usage time.
·No need to adjust the optical axis, easy maintenance
When replacing the deuterium lamp, there is no need to make complicated optical axis adjustments.
·Unique pool structure design achieves high sensitivity
The structure that achieves the highest signal-to-noise ratio and minimizes the influence of refractive index can be easily replaced from the front of the device.
Shimadzu LC-10A Dual Pump Liquid Chromatography System
SPD-10AVPPlus
Wavelength range of 190? 600nm
Noise ± 0.35 × 10-5AU
Drift below ± 2 × 10-4AU/h
Dimensions W260 × H140 × D420mm
Weight 13Kg
Operating temperature range: 4-35 ℃
Shimadzu LC-10A Dual Pump Liquid Chromatography System
Column temperature box CTO-10ASVPPlus
Compact structure, adopting a module temperature control method that can be cooled/heated. The dual temperature control function can maximize the suppression of uneven temperature distribution. Equipped with overheat prevention components and leakage sensors to ensure safe use.
CTO-10ASVPPlus
Temperature control method for semiconductor module
Temperature control range 4? 80℃
Temperature control accuracy ± 0.1 ℃
Dimensions W130 × H415 × D420mm
Weight 12Kg
Operating temperature range: 4-35 ℃
Shimadzu LC-10A Dual Pump Liquid Chromatography System
Chinese Culture LC Workstation LCsolutionLite
Pursue a unified operating environment in the Internet era
Chinese interface, Chinese help menu
Realize LC analysis, report production, and data management with simpler operations
·Create an ideal operating environment
By adopting multiple windows, various operations can be directly accessed, from device control to analysis and data parsing, providing you with an ideal operating environment.
·Easy entry into analysis
By using the assistant bar, data browser, and instrument parameter window for guided analysis, one can quickly access the required operating interface.
Shimadzu LC-10A Dual Pump Liquid Chromatography System
NOS. | DESCRIPTION | Q'TY | UNIT PRICE | AMOUNT | REM | ||
A) | Pump Unit | ||||||
1 | LC-10ATVP Liquid Pump, pump | 2 | set | 5,300.00 | 10,600.00 | ||
2 | Mixer, SUS, mixer | 1 | set | 1,400.00 | |||
B) | Detector detector | ||||||
1 | SPD-10AVP UV-VIS Spectrophotometetric Detector (190-600nm), UV detector | 1 | set | 7,300.00 | |||
C) | Packaged Columns chromatography column | ||||||
1 | Kromstar C18 4.6mm x 15cm, chromatographic column | 1 | pc | 465.00 | |||
D) | Workstion Workstation | ||||||
1 | N2000 Chromatographic Data Workstation | 1 | pc | 200.00 | |||
E) | Sample Injector Injector | ||||||
1 | 8725i Manual Sample Injector (Position Sensing Switch Incorporated) | 1 | set | 1,430.00 | |||
2 | 100ul micro flat head injection needle | 2 | pc | gift | |||