3S-CL-CL2 residual chlorine/total chlorine/ammonium chloride analyzer is an analytical instrument used for online analysis and process monitoring. It uses LED light source for colorimetric analysis and measurementResidual chlorine/total chlorine/ammonium chlorideofConcentration.
In the presence of an appropriate buffer, the chlorine containing sample reacts with N, N-diethyl-phenylenediamine sulfate. Add indicator and buffer, react with chlorine to produce a pink compound. This compound is measured at 525 nanometers. Potassium iodide was added to the measurement of total chlorine and monochloramine.
Product advantages
The instrument adopts a dual compartment housing to ensure complete separation between the electronic components (upper part) and the sampling and analysis components (lower part).
The instrument uses less reagent consumption, reducing operating costs.
The instrument can use an internal dilution module, with a measurement range ranging from trace μ g/L to 1000mg/L; Multiple chemicals can be supplied to meet actual needs.
The analyzer is equipped with a temperature heating module.
The analyzer can be equipped with a refrigerator to keep the chemicals at the appropriate temperature.
The equipment can be set up for automatic calibration, verification, and zeroing, which can reduce downtime and operator intervention, ensuring accurate results.
The analyzer adopts a color touch screen interface to display measurement values and status information. Easy to access menus and functional interfaces. And supports USB data download.
You can choose the dual channel version. By adding an external sequencer, the number of sampling channels can be increased to 4.
The instrument undergoes factory testing before leaving the factory. Simply connect the power supply, sample and reagent pipelines, and the analyzer can operate.
Product Parameters
measurement parameters |
Cl2(ppb, ppm, mg/l). |
size |
604*380*242mm (H*W*D) |
Measuring principle |
Differential colorimetric method DPD colorimetric method (accepted by US Environmental Protection Agency 4500-CIG and ISO7393-2) |
weight |
44 pounds (20kg) |
measurement range |
0.01-2ppm Cl2(26mm reaction tank) 0.02-5ppm Cl2(16mm reaction tank); addInternal dilution, up to 200ppm Cl2 |
power supply |
100-240VAC 50/60Hz standard or 24 VDC Power consumption: MAX80 watts |
通道 |
1,2 adopts built-in switch valve 3,4 use external sequencers |
output |
Two 4-20mA analog signals; One Modbus RTU RS485 digital signal |
frequency analysis |
Freely programmable, close to continuous analysis |
call the police |
2 programmable SPDT passive relays |
cycle time |
2.5 minutes, including pre-treatment before the analysis cycle and cleaning after measurement |
digital input |
Remote start/stop |
reactor |
Temperature heating |
temperature |
41-113℉ (5-45℃) |
water sample |
Temperature: 5-50℃ Connection interface: 6mm Flow rate: 80-500mL/min |
humidity |
10% to 90% non condensing (indoor installation); Outdoor installation requires the use of protective cabinets or rainproof systems |
drainage |
No pressure, standard atmospheric pressure drainage Connection: 12mm (1/2 inch) |
install |
Wall mounted (standard), Desktop support or panel installation (Option) |