Coliynuclear radiationClassification List | |||||
Detectable radiation |
PRODUCT MODEL |
sensor |
range |
γX-ray sensitivity(Cs137,1µSv/h) |
characteristic |
portable | |||||
α、β |
R900 |
ZINC SULFIDE+plastic scintillator |
1Mcps |
-- |
Effective area1702 |
R700 |
ZINC SULFIDE+plastic scintillator |
1Mcps |
-- |
Effective area750px2 |
|
α、β、γ |
R500 |
Geiger counter |
1mSv/h |
7cps |
Large end window cover leather pipe Unable to distinguish the type of radiation |
910 |
Geiger counter |
1mSv/h |
2cps |
Small end window cover leather tube Unable to distinguish the type of radiation |
|
R800 |
CZTsemiconductor |
γ:10mSv/h β:1mSv/h |
7cps |
Simultaneously measuring andDistinguish displayα、βandγray |
|
γ |
RS10 Energy spectrum analyzer |
Φ50x50mmsodium iodide |
1mSv/h |
800cps |
Spectral analysis, energy resolution<> |
10x10x10mm3cadmium zinc telluride |
10mSv/h |
60cps |
Spectral analysis, energy resolution<> |
||
R100 |
GMcounter tube (After energy compensation) |
100mSv/h |
2cps |
persondoseAlarm device, energy range20keV~3MeV |
|
R750 |
Φ80x80mm plastic scintillator |
0.2mSv/h |
2000cps |
Environmental grade, ultra-high sensitivity Energy Range15keV~7MeV(±15%) |
|
RW10 |
φ35×35mmsodium iodide |
1.5mSv/h |
400cps |
100Continuous underwater work |
|
ION-H |
Radiation resistant scintillation sensor |
100Sv/h |
77cps |
High range measurement, optional range10Sv/h、 20Sv/h、50Sv/h、100Sv/h |
|
R700 |
Φ75x75mmplastic scintillator |
0.2mSv/h |
1500cps |
Environmental grade, high sensitivity |
|
R700 |
Φ30x30mmsodium iodide |
1mSv/h |
350cps |
answerγHigh radiation detection efficiency、1Quick response in seconds |
|
|
RC510 γX-ray imaging |
GAGGcrystal |
10mSv/h |
500cps |
High sensitivity sensor, encoding hole imaging technology Quickly identify and locate radiation sources |
|
Neutronsγ |
R60N |
10×10×10(mm3)Koley PatentHType sensor |
10mSv/h(Neutron) |
0.5cps(Cf-252 ) |
A single sensor can simultaneously detect neutrons andγray γPSRR≥1000:1 High sensitivity under the same volume3Hesensor20times |
20×20×20(mm3)Koley PatentHType sensor |
2.7cps(Cf-252 ) |
||||
Wall mountedR2000 | |||||
|
γ
|
GM30A |
Geiger counter |
1mSv/h |
2cps |
Small end window cover leather tube X、γEnergy response:50keV-3MeV |
CS30A |
Φ30x30mm plastic scintillator |
1mSv/h |
350cps |
1Quick response in seconds |
|
CS75 |
Φ75x75mmplastic scintillator |
0.2mSv/h |
1500cps |
Environmental grade High sensitivity |
|
Neutronsγ |
N10 |
10×10×10(mm3)Koley PatentHType sensor |
10mSv/h(neutron) |
0.5cps(Cf-252 ) |
A single sensor can simultaneously detect neutrons andγray γPSRR≥1000:1 High sensitivity under the same volume3Hesensor20times |
N20 |
20×20×20(mm3)Koley PatentHType sensor |
2.7cps(Cf-252 )
|
|||
R900Large area surface pollution meter
sketch
Newly developed by Ke Lei Technology Co., LtdR900Type surface pollution meter,usenewLarge area doubledodgedetectorThe effective detection area is170cm2No need for inflation, greatly reducing the cost of use. Thickness of plastic scintillatoronlydo1mm,effectiveReduced the interference of gamma rays.The thickness of the aluminum film on the incident window is less than3μmYes, that's rightαThe absorption of radiation is only1MeV.R900Can quicklyReliable screening for radiation contamination,Directly detecting the surface of an objectα、βLevel of radioactive surface contamination,α(239Pu)detection efficiencydo30%,βThe detection efficiency is25%.
R900Large area surface pollution meterThe host adopts3.2Inch color touch screen,Upgraded from the traditional handheld watch operating system to the newly developed graphical interface operating system by Koehler, greatly simplifyingThe operational process has been streamlined.Added a patented background removal algorithm, which is more conducive to the accuracy of measurement data. The host provides large capacity storage and data reading functions.Can be connected through the upper computer softwarePCRealize the storage, transmission, and analysis of data,Products available in EuropeCECertification andISO9001Certificate.
application
R900A large-scale surface pollution meter,apply toEnvironmental protection, military, customs, health, nuclear industry, civil aviation, pharmaceuticals, nuclear power plants and other systems.
usea surnamedetectionenvironmentWorkplaceLaboratoryhospital、Isotope production plantTransportation, storage, and inspection of nuclear materials in radioactive environmentsanswerMonitoring of human radioactive contamination levels、Radioactive contamination level of goodssupervisemeasure、Radioactive contamination on surfaces such as workbenches, floors, walls, hands, and clothingMonitoring, etc.
Product characteristics
1. Large measurement window with low measurement limit,High precision
2. Effective detection area170cm2
3. Probe adoptsZnSadd1mmthinplastic scintillatorNo need to inflate
4. Aluminum film covering the entrance window, with a thickness less than3μm
5. Detecting background:αthoroughfare≤ 0.1cps ;βthoroughfare≤ 15cps
6. Detection efficiency:α≥30%(Pu-239);β≥25%
7. manyDisplay mode, capable of displaying simultaneouslyα、βmeasurement result
8. 4500mAhhigh-capacityapproverechargeable lithium battery
9. Provide large capacity data storage function, which can automatically or manually store data
10. Can be connected through the upper computer softwarePCImplement data storage,Transmission and analysis, etc
Technical Parameter
Measure the type of radiation |
α、βsurface radioactivity contamination |
Surface activity response |
αRadiation:40 cps·s-1Bq-1cm2 βRadiation:30cps·s-1Bq-1cm2 |
Detecting effective area |
170cm2 |
sensor |
ZnScoating+thinPlasticScintillation material(The thickness of the plastic scintillator is1mmAluminum film thickness of the incident window<>μm) |
range |
0-1Mcps |
Display units |
cps,Bq/cm2,Bqclass |
detection efficiency |
α≥30%(Yes)Pu-239);β≥25% |
Detecting background |
– αChannel:≤ 0.1cps βChannel:≤ 15cps |
display |
3.2Inch color touchLCDdisplay |
Alarm |
soundlightcall the police,Within the measuring rangeFree setting of alarm values |
accuracy |
±15% |
communication interface |
USB |
Storage function |
Manual or automatic storage |
software |
Real time remote transmission of numerical values to a computer for display, analysis, and recordingAnd download |
operation temperature |
-10℃to+60℃ |
storage temperature |
-20℃to+75℃ |
weight |
host:350gProbe:800g |
power supply |
rechargeable4500mAhlithium battery |
Continuous battery usage time |
>24hour |
Certificate |
ISO9001,European CE |