QLH-500 Beijing Ventilation Aging Test Chamber Manufacturer
Product Introduction:This product meets the requirements of JB/T7444-94 "Air Thermal Aging Test Chamber" and GB/T3512-2001 test standards; Mainly used for heat resistance testing of electrical insulation materials, ventilation aging testing of electronic components and plasticized products.
1、 Main technical indicators
model |
QLH-500 |
||
Internal dimensions: D × W × H |
700×800×900:mm |
||
Dimensions: D × W × H |
1100×920×1800:mm |
||
temperature range |
RT+10℃~200℃ |
RT+10℃~250℃ |
RT+10℃~300℃ |
temperature uniformity |
≤±2℃ |
||
temperature fluctuation |
≤±0.5℃ |
||
Air exchange rate |
1-100 times/h (adjustable) |
||
Adjustable ventilation mode |
Instrument controlled automatic ventilation |
||
Circulation mode |
Fan forced circulation mode |
||
Ventilation time |
1 minute to 99 hours |
||
Time setting range |
0-9999 hours |
||
Power requirements |
AC380 (± 10%) V/50HZ three-phase five wire system |
||
Inner liner material |
SUS304 mirror stainless steel plate |
||
Shell material |
A3 steel plate spray coating |
||
|
Equipment usage conditions: 1. Environmental temperature: 5 ℃ to+28 ℃ (average temperature within 24 hours ≤ 28 ℃) 2. Environmental humidity: ≤ 85% R.H 3. The operating environment requires good indoor ventilation, and no objects should be placed 80 centimeters in front, back, left, right, or right of the machine; | |||
QLH-500 Beijing Ventilation Aging Test Chamber Manufacturer
2、 Control system:
1. The instrument adopts imported dual digit microcomputer PID and At self-tuning high-precision temperature control instrument;
2. Temperature sensor with an accuracy of ± 0.1 ℃, PT100 platinum resistance thermometer with a sampling period of 0.01ms;
3. Ventilation control is automatic instrument control rather than natural ventilation;
4. During automatic ventilation, the instrument sets the ventilation frequency, ventilation time, and stop time to control the driver to open and close the ventilation valve port;
5. The top of the box is equipped with a ventilation outlet, which can be pre adjusted in size;
6. The panel control surface is equipped with ventilation operation display and stop display;
7. Safety protection system: The test chamber is equipped with over temperature, timed, overload protection, and fan overheating protection;
3、 Protection system:
1. Non fuse protection switch
2. Box overheating protection
3. End of experiment indication
4. Overload and leakage protection include automatic shutdown
Service commitment: Free home delivery. After the installation and debugging of the equipment is completed, relevant technical personnel will receive free operation training at the user's site, with no limit on the number of people.