Performance characteristics
·pH(0~100.0)℃automatic/Manual temperature compensation;Dissolved oxygen automatic temperature compensation (0-45) ℃,Automatic salinity compensation (0-45) ppt,Automatic air pressure compensation (80-105) kPa;
·pHAutomatic calibration from 1 to 5 o'clock, with calibration guidance★;
·Self diagnostic function, (incorrect selection of standard solution,pHThe electrode slope exceeds the standard by<85% or>110%,pHThe electrode zero potential exceeds the standard by -60mV or>60mV, etc., and prompts corresponding information; ★
·pHCan be usedCH/ USA/ NISTin all15Calibration of standard solution; Custom standard buffer solution calibration;
·Configure a dedicated dissolved oxygen electrode calibration kit, which automatically calibrates with water saturated air;
·When calibrating salinity, there may be prompts for calibration solution errors or exceeding the standard;
·pHCalibration reminders can be set and calibration time can be viewed;
·Equipped with stable endpoint display and automatic locking display mode, optimized reading accuracy and response time★;
·Compliant with GLP data management practices; Real time clock display, manual, automatic data storage, adjustable timed storage mode, and can display and clear stored values;
·1000 sets of data storage;★
·HavingUSBCommunication interface;Standard data CD; After data transmission to the computer, the measurement serial number, date, time, measurement value, and temperature value can be displayed; ★
·High purity waterpHMeasurement mode ★; Add ammonia pure waterpHMeasurement mode ★;EHMeasurement mode of oxidation-reduction potential★;
·Temperature units can be selected℃and℉;
·The instrument can be reset to factory settings with just one click★;
·Automatic shutdown, low voltage display;
·High quality three in one configurationpHElectrode, Swiss pure imported high-performance glass compositepHElectrodes,The dissolved oxygen electrode is equipped with a temperature sensor and a salinity sensor;
·Randomly give away three dissolved oxygen diaphragm caps;★
·360 degree steering and vertical lifting universal electrode bracket;★
·90 × 68mm large screen LCD, large font display, white backlight, clear, easy to operate;
·The operation interface is simple, with touch sensitive panel buttons that are moisture-proof, dustproof, and oil resistant, and have a long service life;
·Compliant withIP57Dustproof and splash proof rating, can be used in1Enduring underwater in meters deep water30Minute test;
Technical Parameter
Instrument level |
0.001level |
|
measuring rangeMeasuring range |
pH |
(-2.000~19.999) pH |
mV |
(±1999.9)mV |
|
dissolved oxygenDissolved oxygen |
(0~20.00)mg/L(ppm) |
|
Dissolved oxygen saturationDissolved oxygen saturation |
(0.0~200.0)% |
|
temperature |
(-10~110)℃ |
|
|
resolutionResolution |
pH |
0.001/0.01/0.1 pH |
mV |
0.1 mV |
|
dissolved oxygenDissolved oxygen |
0.01/0.1 mg/L |
|
Dissolved oxygen saturationDissolved oxygen saturation |
0.1/1% |
|
temperatureTemperature |
0.1℃ |
|
Basic errorBasic error |
pH |
±0.002 pH |
mV |
±0.03%FS |
|
dissolved oxygenDissolved oxygen |
±0.30mg/L |
|
temperatureTemperature |
±0.4℃ |
|
response timeResponse time |
30s (25 ℃, 90% response) |
|
Residual currentResidual current |
0.10mg/L |
|
Standard configuration
PD5200host,S2211TBHigh quality plastic shell triple compositepHElectrodes,S3221glasspHComposite electrode,T2132Temperature electrode,G1218Dissolved oxygen electrode(1.5m),ST901Intelligent mixer, universal electrode holder,G1218-1Dissolved oxygen electrode diaphragm cap,pHStandard buffer solution,G1218-3The solution inside the dissolved oxygen electrode,USBCommunication cable, communication software CD,9VThe power adapter.
PC5200 host, S2211T molded pH3-in-1 combination electrode, S3221 glass pH combination electrode, T2132temperature electrode, G1218 DO electrode, ST901 intelligent mixer, universalelectrode stand, G1218-1 DO electrode diaphragm cap, pH standard buffersolution, G1218-3 DO electrode solution, USB communications cable,communications software CD, 9V power adapter.
Optional electrode
·S3221DglasspHComposite electrode——Used for high-purity waterpHMeasurement.
S3221D glass pH combinationelectrode - for pH measurement of highly pure water
·S2211FplanepHComposite electrode——Used for wetting solid or semi-solid surfacespHMeasurement, such as skin, paper, fabric, meat, and fruits; And colloidal solutions and trace solutionspHMeasurement.
S2211F flat pHcombination electrode - for the pH measurement of solid or semi-solid wetsurfaces (e.g. skin, paper, cloth, meat and fruits); colloidal solutio