Performance characteristics
·pH(0~100.0)℃automatic/Manual temperature compensation; conductivity(0-50.0) ℃ automatic/manual temperature compensation;
·pHAutomatic calibration from one to five o'clock, conductivityAutomatic calibration from one to four o'clock, with calibration guidance function;
·Self diagnostic function, (incorrect selection of standard solution,pHThe electrode slope exceeds the standard by<85% or>110%, and the pH electrode zero potential exceeds the standard by -60mV or>60mV, etc., and corresponding information is prompted; ★
·Standard solution exceeding the standard and pollution alarm prompt; ★
·pHCan be usedCH/ USA/ NISTin all15Calibration of standard solution; conductivityCan be calibrated using a total of 8 standard solutions from CH/USA;Custom standard buffer solution calibration;
·Calibration reminders can be set and calibration time can be viewed;★
·Equipped with stable endpoint display and automatic locking display mode, optimized reading accuracy and response time★;
·Compliant with GLP data management practices; Real time clock display, manual, automatic data storage, adjustable timed storage mode, and can display and clear stored values;
·1000 sets of data storage;★
·HavingUSBCommunication interface;Standard data CD; After data transmission to the computer, the measurement serial number, date, time, measurement value, and temperature value can be displayed; ★
·High purity waterpHMeasurement mode★Add ammonia to pure waterpHMeasurement mode★;EHMeasurement mode of oxidation-reduction potential★;
·Switchable conductivity TDS、 Salinity and resistivity measurement modes;
·Ultra pure water (ultra-low conductivity) stable measurement mode;★
·ConductivityElectrode constant selection; Temperature compensation coefficient setting; Reference temperature selection ★;
·Temperature units can be selected℃and℉;
·The instrument can be reset to factory settings with just one click★;
·Automatic shutdown, low voltage display;
·High quality three in one configurationpHElectrode, Swiss pure imported high-performance glass compositepHElectrode andConductive electrode with temperature compensation ★
·360 degree steering and vertical lifting universal electrode bracket;★
·90 × 68mm large screen LCD, large font display, white backlight, clear, easy to operate;
·The operation interface is simple, with touch sensitive panel buttons that are moisture-proof, dustproof, and oil resistant, and have a long service life;
·Compliant withIP57Dustproof and splash proof rating, can be used in1Enduring underwater in meters deep water30Minute test;
Technical Parameter
Instrument level |
PH level 0.001, conductivity meter level 0.5 |
|
measuring rangeMeasuring range |
pH |
(-2.000~19.999) pH |
mV |
(±1999.9)mV |
|
conductivityConductivity |
0.00 μ S/cm to 2000mS/cm includes: (0.00~20.00)μS/cm; (20.0~200.0)μS/cm; (200~2000)μS/cm; (2.00~20.00)mS/cm; (20.0~200.0mS/cm;(200~2000)mS/cm |
|
resistivityResistivity |
(0~100)MΩ•cm |
|
TDS |
(0~100)g/L |
|
salinitySalinity |
(0~100)ppt |
|
temperatureTemperature |
(-10~110)℃ |
|
resolutionResolution |
pH |
0.001/0.01/0.1 pH |
mV |
0.1 mV |
|
conductivityConductivity |
0.01/0.1/1μS/cm; 0.01/0.1/1mS/cm |
|
temperatureTemperature |
0.1℃ |
|
Basic errorBasic error |
pH |
±0.002 pH |
mV |
±0.03%FS |
|
conductivityConductivity |
±0.5%FS |
|
temperatureTemperature |
±0.4℃ |
|
Standard configuration
PC5200 host, S2211TB high-quality plastic shell triple composite pH electrode, S3221 glass pH composite electrode, T2132 temperature electrode, E2224T glass conductivity electrode (K=1, ATC), universal electrode holder, pH standard buffer solution, conductivity correction solution, USB communication cable, communication software CD, 9V power adapter.
PC5200host, S2211T molded pH 3-in-1 electrode, S3221 glass pH combination electrode,T2132 temperature electrode, E2224T glass conductivity electrode (k=1,ATC),universal electrode stand, pH standard buffer solution, conductivitycalibration solution, USB communications cable, communications software CD, 9Vpower adapter.
Optional electrode
·S3221DglasspHComposite electrode–Used for high-purity waterpHMeasurement;
S3221D glass pH combination electrode – for pH measurement of highly purewater;
S2211FplanepHComposite electrode–Used for wetting solid or semi-solid surfacespHMeasurement, such as skin, paper, fabric, meat, and fruits; And colloidal solutions and trace solutionspHMeasurement.
S2211F flat pH combination electrode - for the pH measurement of solid orsemi-solid wet surfaces (e.g. skin, paper, cloth, meat and fruits); colloidalsolutions and micro solutions
E3224TType of conductive electrode(k=0.1, ATC)Used for measuring the conductivity of high-purity water;
E3224T conductivity electrode (k=0.1, ATC) - for conductivity measurement ofhighly pure water;
E1214TType of conductive electrode(k=10, ATC)-Used for high concentration conductivity measurement.