Application fields:
Instruments are used for measuring disinfectants in drinking water, swimming pools, industrial cooling water, water plant effluent, and municipal pipeline monitoring
Instrument features:
· The aluminum IP66 instrument transmitter, overflow device flow tank, and electrode are installed on a PVC panel
· Factory tested and ready to use
· Self cleaning three electrode system (platinum electrode for working electrode and reverse electrode)
· Input and output overvoltage protection
· Two channels of 0/4-20 mA analog signal output can be used for disinfectant or temperature output or continuous control output
· Integrated alarm dry node programmable alarm values and instrument fault alarms
· Electrode based online disinfectant analyzer, such as hypochlorous acid, residual chlorine, chlorine dioxide, ozone, bromine, and iodine
· The pH electrode can achieve instant pH compensation during residual chlorine measurement (option)
· No reagents, no consumables
· Factory tested, ready to use upon installation
· PH electrode with cable
Technical Specifications:
· Range:
-Ozone: 0.000 to 1.000 mg/l
-Hypochloric acid, residual chlorine: 0.00 to 5.00 mg/l
-Chlorine dioxide, bromine, iodine: 0.00 to 3.00 mg/l
· connect:
-Inlet hose Ø 8 x 6 mm
-Outlet hose Ø 20 x 15 mm (1/2 ')
Options:
· PH or ORP (Redox) electrode with cable
· Communication interface