Mettler pH electrode InPro 2000 250/PT100/9816
Specification - pH Sensor InPro 2000 250/PT100/9816
Simulation or ISM | Analog |
Measurement parameters | pH |
PH glass film type | High Alkali glass (HA) |
Single liquid connection or double liquid connection | Single Junction Refillable |
sensor length | 250 mm |
temperature signal | PT100 |
Type of connector | VP |
Solution grounding/redox metal | None |
electrolyte | Viscolyt |
pH range | 0-14 pH |
Pressure (bar) | 0 to 6 barg (in presurized housing) |
Pressure (psi) | 0 to 87 psig (in presurized housing) |
reference system | Argenthal with silver-ion trap |
disinfectant | Yes |
High voltage resistance | Yes |
membrane | Ceramic |
ATEX certificate | SEV 14 ATEX 0168 X<br> Ex ia IIC T6/T5/T4/T3 Ga/Gb |
FM Approval | IS/ I, II, III /1/ABCDEFG /T6 |
IECEx certificate | IECEx SEV 14.0025X<br> Ex ia IIC T6/T5/T4/T3 Ga/Gb |
Measurement temperature range (imp) | 32 to 284° F |
Measurement temperature range (index) | 0 to 140° C |
Material number (s) | 52001428 |
Mettler pH electrode InPro 2000 250/PT100/9816
9848: Suitable for biochemistry, media containing proteins, organic solvents, and low temperatures
9816: Suitable for chemical processes, slow electrolyte flow rate 9830: Suitable for media containing a large amount of organic solvents 9823: Classic electrolyte, fast electrolyte flow rate, helps with membrane cleaning
6.7 How to eliminate alarms
According to the alarm information prompt, adjust the alarm parameters to the normal range and the exclamation mark will automatically disappear. Replace the electrode and recalibrate.
After testing for a period of time, the pH of 6.8M420 suddenly became abnormal. After restarting, it returned to normal mainly due to an error in the wiring of the analog signal electrode. It is necessary to strictly follow the instructions for wiring Pay special attention to the annotations and explanations.
Difference in mouth model identification
For example: InPro4260SG/120/PT1000&InPro4260/120/PT1000.
So the mouth When wiring electrodes with SG function, connect the SG terminal to a blue cable: electrodes without SG function need to be short circuited to the reference and SG terminals.
During the calibration process, the electrode response is slow/sloping, and the zero offset is large. pH electrodes can age due to differences in production and usage environments (temperature, pH). If the electrode is calibrated multiple times with fresh buffer solution, this phenomenon indicates that the electrode's lifespan has expired.
Of course, the mouth. Improper operation can also cause this phenomenon. Check and avoid the following situations:
·Is the electrolyte inside the sensitive membrane full? If not, it needs to be vigorously shaken a few times to ensure that the electrolyte inside the sensitive membrane exceeds 1/2
·The electrode in the beaker should be suspended as much as possible
·The electrode sensitive membrane or diaphragm has attachments You can contact our company