Product Application
■ Fuel boiler
■ Plastic industry
Drying machinery
■ Cogeneration unit
Industries such as pharmaceuticals, food, electricity, metallurgy, glass, chemicals, rubber, and textiles
Technical data
| Thermocouple core | Nickel chromium nickel silicon (K): -200~800 ℃ |
| Nickel chromium copper nickel (E): -200~800 ℃ | |
| Iron copper nickel (J): -200~750 ℃ | |
| Copper Copper Nickel (T): -200~350 ℃ | |
| accuracy class | Level II |
| Extension tube | Stainless steel, length 50mm |
| Minimum depth of placement | ≥50mm, Excluding special products |
| process connection | Stainless steel, threaded/chuck/movable flange |
| Protection tube | Stainless steel<800 ℃, diameter: 6mm, 8mm, 10mm |
| response time | t0.5,5s~90s |
| Junction box | Cast aluminum casing, outlet thread M16 * 1.5, IP65 |
Instructions
·The standard material for protective tubes, extension tubes, and process connections is stainless steel 1.4301. For other stainless steel materials, please consult our sales staff and indicate them when selecting;
·For special ranges or the use of other thermocouple inserts such as B, R, S, N division marks, please consult our sales personnel and indicate them when selecting;
·The specific parameters of metal and non-metal protective tube materials are detailed in Table 7600-2 on page 91 and Table 7600-3 on page 92;
·When choosing a module with a temperature transmitter, please consult our sales personnel.