Product Name:
Intelligent integrated vortex flowmeter -
model:
LWG
Product Summary:
Intelligent integrated vortex flowmeter is a volumetric flowmeter that measures the volumetric flow rate, standard condition volumetric flow rate, or mass flow rate of gases, vapors, or liquids based on the Karman vortex principle. And it can be used as a flow transmitter for vortex flow meters in automation control systems.1、 Overview
Intelligent integrated vortex flowmeter is a volumetric flowmeter that measures the volumetric flow rate, standard condition volumetric flow rate, or mass flow rate of gases, vapors, or liquids based on the Karman vortex principle. And it can be used as a flow transmitter for vortex flow meters in automation control systems.
This instrument adopts advanced differential technology, combined with isolation, shielding, filtering and other measures, to overcome the problems of poor seismic resistance and small signal data disorder of similar products. It also adopts unique sensor packaging technology and protective measures to ensure the reliability of the product. There are two types of products: basic type and composite type. The basic type measures a single flow signal; The composite type can simultaneously measure temperature, pressure, and flow rate. Each type has a whole or split structure to adapt to different installation environments.
2、 Characteristics and Applications of Instruments
characteristic:
No movable parts, long-term stability, simple structure for easy installation and maintenance
Adopting anti-interference circuit and anti vibration sensing head, it has certain anti environmental vibration performance
Using ultra-low power single-chip microcomputer technology, one 3.2V10AH lithium battery can be used for more than 5 years
Correction of instrument coefficient nonlinearity by software to improve measurement accuracy
Low pressure loss and wide range of measurement
Using EEPROM for power down protection of accumulated traffic, with a protection time greater than 10 years
Purpose:
This instrument can be widely used for measuring the flow of water supply and drainage, industrial circulation, sewage treatment, oil and chemical reagents, compressed air, saturated and superheated steam, natural gas, and various media in large, medium, and small pipelines.
3、 Technical parameters
Nominal meridian (mm) |
15,20,25,40,50,65,80,100,125,150,200,250,300 |
Instrument material |
1Cr18Ni9Ti |
Nominal pressure (Mpa) |
PN1.6Mpa; PN2.5Mpa; PN4.0Mpa |
Temperature of the tested medium (℃) |
-40~+250℃; -40~+350℃ |
environmental condition |
Temperature -10~+55 ℃, relative humidity 5%~90%, atmospheric pressure 86~106Kpa |
accuracy class |
Measuring liquids: ± 0.5 of the indicated value |
Measuring gases or vapors: |
± 1.0 and ± 1.5 of the indicated value |
Range ratio |
1:10; 1:15 |
Resistance loss coefficient |
Cd<2.6 |
output signal |
Optional 4-20mA or pulse |
power supply |
Sensor:+12VDC,+24VDC (optional) Transmitter:+24VDC On site display type: The instrument comes with a 3.2V lithium battery |
Signal transmission line |
STVPV3 × 0.3 (three wire system), 2 × 0.3 (two wire system) |
Installation method |
Flange clamping or integrated flange connection |
Distance |
≤500m |
Signal line interface |
Internal thread M20 × 1.5 |
Explosion proof grade |
ExdIIBT6 |
protection grade |
IP65 |
Allow vibration acceleration |
1.0g |
4、 Measurement Range Table
Caliber (mm) |
Liquid measurement range (m3/h) |
Gas measurement range (m3/h) |
DN15 |
1.2-6.2 |
5-25 |
DN20 |
1.5-10 |
8-50 |
DN25 |
1.6-16 |
10-70 |
DN32 |
1.9-19 |
15-150 |
DN40 |
2.5-26 |
22-220 |
DN50 |
3.5-38 |
36-320 |
DN65 |
6.2-65 |
50-480 |
DN80 |
10-100 |
70-640 |
DN100 |
15-150 |
130-1100 |
DN125 |
25-250 |
200-1700 |
DN150 |
36-380 |
280-2240 |
DN200 |
62-650 |
580-4960 |
DN250 |
140-1400 |
970-8000 |
DN300 |
200-2000 |
1380-11000 |
5、 Xi'an Yunyi Intelligent Integrated Vortex Flow Meter Selection Table
model |
|
|
|
|
|
|
|
|
LWG |
Pipeline type |
|
|
|
|
|
|
|
LWC |
plug-in |
|
|
|
|
|
|
|
|
|
pipe diameter |
|
|
|
|
|
|
|
|
20-300mm |
|
|
|
|
|
|
|
|
|
medium |
|
|
|
|
|
|
|
|
1 |
gas |
|
|
|
|
|
|
|
2 |
compressed air |
|
|
|
|
|
|
|
3 |
liquid |
|
|
|
|
|
|
|
4 |
Saturated steam |
|
|
|
|
|
|
|
5 |
superheated steam |
|
|
|
|
|
|
|
|
output signal |
|
|
|
|
|
|
|
|
0 |
Voltage pulse output |
|
|
|
|
|
|
|
1 |
4-20mA output |
|
|
|
|
|
|
|
2 |
Battery powered, on-site LCD display |
|
|
|
|
|
|
|
3 |
4-20mA output, on-site LCD display |
|
|
|
|
|
|
|
|
explosion-proof |
|
|
|
|
|
|
|
|
B |
Intrinsic safety explosion-proof |
|
|
|
|
|
|
|
not have |
|
|
|
|
|
|
|
|
|
accuracy |
|
|
|
|
|
|
|
|
0.5 |
Level 0.5 |
|
|
|
|
|
|
|
1 |
Level 1 |
|
|
|
|
|
|
|
|
Connection method |
|
|
|
|
|
|
|
|
1 |
Flange connection |
|
|
|
|
|
|
|
2 |
Flange card installation |
For example, LWG-50-4-3-0.5-2 refers to a flange mounted flowmeter with a diameter of 50mm, a current output LCD display, and a precision level 1 vortex flowmeter.
6、 Dimensional drawing of flange card installation and flange connection vortex flowmeter sensor
7、 Comparison Table of Sensor Sizes for Vortex Flow Meter
explain:
When measuring steam, the total height of the sensor is H+100mm;
Card mounted sensors must be equipped with dedicated 2.5MPa or 4.0MPa concave flanges;
The outer diameter D of the flange type is the outer diameter of the flange with a nominal pressure of 2.5 MPa, which is matched with the flange according to the JB80-51 standard.
8、 Installation diagram of flange card installation and flange connection vortex flowmeter