VIP member
High power standard 2600A series 250W/2000W
Product Description: Based on thermistor, DC substitution method design can be directly traced back to high-power calorimeter calibration. Through typ
Product details
Product Description:
Design of DC Substitution Method Based on Thermistor
• Directly traceable to high-power calorimeter
Calibration of pass through and terminal power probes: Bird, R&S, Anritsu, Keysight
• Frequency range from 10 kHz to 3 GHz
• Power of 250 W (2601A and 2602A), 2000W (2605A and 2606A)
• Uncertainty<2.0% (2601A and 2602A),<1.0% (2605A and 2606A)
• Compatible with TEGAM1830A power controller
2601A: High power transfer standard 250W from 250 kHz to 1 GHz
2602A: High power transfer standard from 1 GHz to 3 GHz, 250W2605A: High power transfer standard from 10 kHz to 100 MHz 2000W
2606A: High power transfer standard 2000W from 50 MHz to 500 MHzThe TEGAM coaxial high-power transmission standard 2600A series uses the classic "DC substitution method based on thermistor" to achieve direct traceability of high power, with a frequency range covering 10KHz to 3GHz and a power range of 250W or 2000W.
The TEGAM high-power transmission standard 2600A series consists of a precision directional coupler and a thermistor holder. The additional large heat sink can minimize the performance deviation caused by temperature changes, making the standard's performance highly accurate and stable; It is an ideal device for calibrating other medium and high power calibrated probes (such as Bird, RS, Anritsu, and Keysight) as a power transfer standard.
The 2600A series can be traced back to high-power calorimeters, thereby achieving microwave/RF high-power traceability to International Standard Units (SI); The calibration of the 2600A series can be conducted at NIST or other equivalent metrology institutions.
2600A series high power transmission standard compatible with TEGAM 1830A power meter/power controller; As part of the TEGAM microwave/RF low-power calibration system, 1830A has been widely used by numerous calibration laboratories.
Similar to the TEGAM low-power transmission standard, the 2600A series adopts a pass through design that complies with calibration regulations and the methods and requirements of the "standard transmission method" in practice.
The 2600A series uses N-type coaxial connectors.
Online inquiry