HQ06 digital leakage current measuring instrument calibrator
HQ06 digital leakage current measuring instrument calibrator
Product details
HQ06 digital leakage current measuring instrument calibrator

1. Overview
The HQ06A digital leakage current tester calibrator (hereinafter referred to as the calibrator) is designed and manufactured in accordance with relevant national regulations, with a five digit display, suitable for the calibration of newly manufactured, used, and repaired leakage current testers (passive instruments). This calibrator is easy to operate, easy to read, has high accuracy, small size, and is widely applicable to various levels of metrology departments.
2. Technical performance
Project | Range | Output range | maximum output | accuracy |
ACV | 300V | 0~300.00V | 50mA | 0.1%+0.05% full scale |
ACA | 200μA | 0~200.00μA | 0.5V | |
2mA | 0~2.0000mA | 5V | ||
20mA | 0~20.000mA | 25V | ||
200mA | 0~200.00mA | 25V | 0.1%+0.05% full scale | |
DCA | 200μA | 0~200.00μA | 0.5V | 0.1%+0.05% full scale |
2mA | 0~2.0000mA | 5V | ||
20mA | 0~20.000mA | 25V | ||
200mA | 0~200.00mA | 25V | 0.1%+0.05% full scale | |
f | 50Hz、60Hz、400Hz、 | 1% |
Communication distortion: not greater than 1%.
Ripple factor: not greater than 1%.
Stability: not greater than 0.05%/2 minutes of full scale.
HQ06B digital leakage current measuring instrument calibrator
1. Overview:
The HQ06B leakage current tester calibrator is designed and manufactured in accordance with relevant national regulations. Applicable for the calibration of leakage current testers (active instruments) for new manufacturing, use, and repair. This calibration instrument is small in size, easy to operate, and convenient to use, making it the preferred measuring instrument for all levels of metrology departments.
2. Technical parameters
Project | Range | measuring range | accuracy |
ACA | 200μA | 0~200.0μA | 0.1% reading+0.05% full scale |
2mA | 0~2.000mA | ||
20mA | 0~20.00mA | ||
200mA | 0~200.0mA | 0.1% reading+0.05% full scale |
Gear setting: 7-speed resistor switch
Resistance range: approximately 471 Ω~10.00047M Ω (21 ℃+2 ℃, humidity less than 50%)
Current detection range: 200 μ A to 200mA; Accuracy: 0.1% of reading+0.05% of full scale (21 ℃+2 ℃, humidity less than 50%; not exceeding 200 μ A)
Assessment accuracy, for reference only)
HQ06 digital leakage current measuring instrument calibrator This article has been read 3145 times