HIOKI3506-10 Japanese capacitance testerBasic parameters
Measurement parameters C (capacitor), D (loss coefficient tan δ), Q (1/tan δ)
measuring range C:0.001fF~15.0000μF,D: 0.00001 ~ 1.99999,Q:0.0 ~ 19999.9
Basic accuracy (Representative value) C: ± 0.14% rdg. D: ± 0.0013
Measurement Frequency 1kHz, 1MHz
Measure signal level 500mV, 1V rms
output resistance 1 Ω (range above 2.2 μ F at 1kHz), 20 Ω (range other than the above)
display LED (6-digit display, full range determined by the effective distance of the points)
Time 1.5ms:1MHz, 2.0 ms:1kHz
function BIN classification measurement, trigger synchronous output, measurement condition memory, measurement value comparison function, average value function, Low-C filtering function, vibration function, current detection and monitoring function, output voltage value monitoring function, control input/output (EXT. I/O), RS-232C interface, GP-IB interface
power supply AC 100/120/220/240V ± 10% (optional), 50/60Hz, maximum 40VA
Volume and weight 260W × 100H × 298Dmm, 4.8kg
attachment Power cord x 1, power backup fuse x 1, user manual x 1
HIOKI3506-10 Japanese capacitance testerTesting tools or cables
Corresponding to 1MHz testing, low capacity, high-precision, high-speed testing
High speed measurement with a simulated measurement time of 0.6ms (1MHz)
Improved anti-interference ability and high repeatability accuracy can also be achieved on the production line
Under 1kHz and 1MHz measurement, low capacitance surface mount can be stably measured
Distinguish capacity based on BIN measurement
Input with probe/test fixture, physical not included. According to the measurement purpose, different models have different probe/test fixture options. RS-232C can be connected using crossover cables available on the market that support network connectivity. The RS-232C connection line 9637 can be used except for hardware flow control.
4-terminal probe L2000, DC~8MHz, 1m long
4-terminal Kelvin clamp 9140-10, DC~200kH, 50 Ω, 1 m long
Tweeze shaped probe L2001, wire length 73cm, DC~8MHz, 50 Ω, front electrode spacing: 0.3-6mm (IM9901: JIS size 1608-5750) (IM9902: JIS size 0603-5750)
Test fixture 9261-10, wire length 1m, DC~5MHz, characteristic impedance 50 Ω, measurable port diameter: 0.3~1.5mm
Test fixture 9262, DC~8 MHz, Direct connection type
SMD test fixture 9263, direct connection type, DC~5 MHz test size: 1mm~10.0mm
SMD test fixture 9677, used for SMD DC~120MHz with electrodes on the side, test sample size: 3.5mm ± 0.5mm
SMD test fixture 9699, used for SMD DC~120MHz with electrodes at the bottom, test sample size: width 1.0~4.0mm, height below 1.5mm