Gas bath constant temperature cultivation shaker
Adopting microcomputer control technology to ensure rapid reaching of the set temperature and high temperature control accuracy. The temperature, spee
Product details

Detailed information on HZ-9211KB gas bath constant temperature cultivation shaker:



The HZ-9211KB air constant temperature oscillator is mainly suitable for oscillation cultivation of various liquid and solid compounds such as biology, biochemistry, cells, and bacterial strains in scientific research departments such as universities, medical institutions, petrochemicals, health and epidemic prevention, and environmental testing. This machine has the characteristics of reasonable structure, simple operation, and high stability performance, making it an ideal equipment for laboratory workers to handle with ease.

1、 Main technical parameters:

1. Power supply: 220V 50Hz

2. Heating power: 300W

3. Timing range: 0-99.9 hours

4. Oscillation frequency: 0-300 revolutions per minute

5. Oscillation amplitude: 30mm

6. Constant temperature range: room temperature~60 ℃

7. Tray size: 420 × 330mm, spring mesh clamp.

2、 Usage:

Firstly, clamp the sample to be cultured by oscillation onto the tray fixture, turn on the power supply, and turn on the power switch indicator light to indicate that the internal power supply has been connected. Turn on the oscillation switch and set the required oscillation frequency and time according to the GTCS-2013 DC motor speed controller. Example: Set the oscillator to work 200 times per minute for 1 hour and 30 minutes before stopping.

1: Close the power switch, and the upper row will display the original speed value. (Last running speed) The original timing value is displayed on the right

2: Press the "" (speed) key to set the speed to 200.

3: Press the H (hour) key to set the timed hour to 1.

4: Press the M (minute) key to set the timer minute to 30. (Set 0:00 to indicate continuous operation in a constantly open state at irregular intervals)

Press the RUN button to start the machine slowly until it reaches 200 revolutions per minute, and start at the same time. Then run steadily for 1 hour and 30 minutes before stopping and sounding an alarm for 10 seconds. Indicates that the scheduled run has been completed.

6: Press stop to stop running at any time (the timer also ends at this time).

7: Green is the running light, and red is the stop light. The corresponding indicator light is on, indicating that the controller is in the corresponding state.

If heating is required, turn on the heating switch, press the "SET" key or "∧", "∨" to enter the temperature setting state. The lower display shows "SP", and the upper display shows "temperature setting value". Then press the "∧" or "∨" keys to increase or decrease the temperature setting value to the set working temperature. After setting, press the "SET" button to acknowledge and exit the setting state.

For detailed operation, please use according to the temperature control table manual.

3、 Attention:

1. The instrument casing should be properly grounded to prevent accidents.

2. It is strictly prohibited for various solutions to enter the studio to avoid damaging the host.