| GXH-3011A portable infrared CO analyzer 0-50.0 × 10-6 pump suction infrared carbon monoxide analyzerDetailed information: |
|
1、 Overview
This instrument complies with the national standards of GB/T18204.23-2000 "Method for the determination of carbon monoxide in the air of public places" and GB/T9801-88 "Determination of carbon monoxide in air quality - Non dispersive infrared method"; Comply with the national metrological verification regulations of JJG635-1999 "Infrared Gas Analyzer for Carbon Monoxide and Carbon Dioxide". The main technical indicators of this instrument meet the technical requirements of national second-class instruments and can obtain a second-class calibration certificate from the Chinese Academy of Metrology. This instrument has a built-in zero adjustment filter, a six way valve for switching zero adjustment and measurement, and is easy and flexible to operate.
2、 Main technical indicators ·Measurement principle: Non dispersive infrared gas analysis method/Non dispersive infrared method (NDIR) ·Sampling method: built-in pump suction ·Measurement range:0~50.0×10-6Or 0~200.0×10-6 ·Resolution:0.1×10-6 ·Linearity: ≤± 2% F · S ·Repeatability: ≤ 1% F · S ·Zero drift: ≤± 2% F · S/4h ·Span drift: ≤±2% F·S/4h ·Start time:≤30min ·Response time:≤30S ·Extraction flow rate: ≤ 1.5L/min ·Standard configuration: host, sampler, strap, rechargeable battery, charger, small screwdriver, technical documents, aluminum alloy box ·Power supply: AC/DC dual-use,220VAC or internal battery ·Dimensions: 245mm × 190mm × 85mm ·Weight: ≤ 3kg |

GXH-3011A Portable Infrared CO Analyzer