GB/T5137.3-2002 Radiation Resistance Tester/Radiation Resistance Test Chamber
Product Usage:
This product is suitable for UV resistance testing of various laminated glass materials such as automotive laminated glass, architectural laminated glass, and fire-resistant glass. It is widely used in deep processing glass enterprises and quality supervision and inspection departments at all levels. The radiation resistance detector is a specialized testing equipment designed and manufactured in accordance with the national standard GB/T5137.3-2002 of the People's Republic of China, "Test Methods for Radiation, High Temperature, Humidity, Combustion, and Simulated Climate Resistance of Automotive Safety Glass"
1、 Technical indicators
| PRODUCT MODEL | ZN-F |
| Internal dimensions (D * W * H) mm | 600×500×800 |
| Dimensions (D * W * H) mm | 850×690×1200 |
| Irradiation light source | Ozone free quartz tube medium pressure mercury vapor arc lamp tube length: 360mm, tube diameter: 9.5mm, arc length: 300 ± 14mm |
| Power supply and transformer power supply | 220cV, transformer: working power: 750 ± 50W, starting peak voltage: 1100V, working voltage: 500 ± 50V |
| Sample size and thickness | Size: 76mm × 300mm, thickness: 6mm~30mm |
| Sample barrel rotation speed | 1-5 revolutions per minute |
| Temperature Control Range | Room temperature~100 ℃ |
| Time setting range | 0~300h, accuracy: 0.1s |
GB/T5137.3-2002 Radiation Resistance Tester/Radiation Resistance Test Chamber
2、 Product Features
1. This instrument uses an ozone free quartz tube type medium pressure mercury vapor arc lamp as the radiation source to irradiate laminated glass, in order to determine whether there will be significant discoloration or a decrease in transmittance of the glass after a period of irradiation.
2. This instrument consists of a power supply, ultraviolet lamp, sample tank, motor, reducer, control circuit, etc. The ultraviolet lamp is vertically placed on the vertical axis of the sample barrel inside the instrument. The sample barrel is driven by a motor and a reducer to rotate around the vertical axis. The sample is placed on the inner wall of the sample barrel to obtain uniform radiation illumination through the rotation of the sample barrel. The control circuit controls the irradiation time and the temperature inside the instrument cavity to meet the test requirements.
3、 Operation method
1. Clean the sample, first test the transmittance of the sample according to the GB5137.2 method, then place the sample into the sample barrel of this testing instrument and close the sample compartment door.
2. Adjust the time and temperature regulators on the control panel together, set the irradiation time and sample test temperature.
3. Connect the instrument power supply, press the "Power" button, and turn on the main power switch of the instrument.
4. Press the "UV Lamp" button to turn on the UV lamp tube.
5. Press the "Sample Bucket" button to start rotating the sample bucket.
6. Press the "Temperature" button to activate the temperature control function. When the temperature inside the chamber exceeds the set temperature, the axial flow fan at the top of the instrument automatically starts to dissipate heat, in order to achieve constant temperature.
7. Press the "timer" button to start timing. When the running time reaches the preset time, all work will be stopped, and only the main power supply will be kept in the on standby state.
8. Press the "Power" button to turn off the instrument.
9. Open the sample compartment door, take out the sample, and test the transmittance of the irradiated sample
4、 Precautions
1. Due to the high power of the ultraviolet lamp, the power transformer of this instrument will generate high temperatures. The machine has been designed with an air duct for the transformer to dissipate heat. Therefore, when the instrument is placed indoors, it should be kept 20cm away from the wall for heat dissipation.
2. Transformers have magnetic saturation characteristics, so it takes a few minutes to turn off the UV lamp and then turn it back on, otherwise the UV lamp cannot be turned on.
5、 After sales service:
One year warranty, lifetime maintenance, regular follow-up visits, listening to customer suggestions, and establishing customer service records;
We provide lifelong free and usage guidance services. Our company promises to respond to customers within 1 hour. If technical service personnel need to be sent to the site for resolution, we promise to send personnel within 24 hours (except for special circumstances such as statutory holidays)