The Hongji Clean Y09-6H3 Laser Dust Particle Counter (handheld, 2.83L/min) is a detection instrument used to measure and analyze the size and quantity of dust particles per unit volume of air in clean environments, and to determine the cleanliness level of clean environments according to standards.
|
|
Particles in the air scatter when exposed to light, a phenomenon known as light scattering. Light scattering is related to factors such as particle size, wavelength of light waves, refractive index of particles, and their light absorption characteristics. However, in terms of scattered light intensity and particle size, there is a basic rule that the intensity of scattered light from particles increases with the increase of their surface area. In this way, by measuring the intensity of scattered light, the size of particles can be inferred, which is the basic principle of light scattering particle counters. |
|
|
Y09-6H3 can be operated in manual mode using the six keys on the instrument control panel. The measurement data can be saved in the instrument's memory, which can hold approximately 256 sets of data. The saved data can be printed using a random printer or sent to a computer (special software needs to be purchased). |
Sampling volume of 2.83 liters per minute
Lithium battery can be fully charged and used sustainably for more than 6 hours
The aviation aluminum shell is not prone to dust accumulation, easy to clean, and has a beautiful appearance
3 particle size channels: 0.3, 0.5, 5.0 μ m
LCD screen display, simultaneous measurement and display of various particle sizes
Real time printing of data, the machine can store 256 pages of data
Optional software for computer connection
Handheld design, convenient to carry and measure at any time
| Overall dimensions:95*160*45mm |
|
Display mode:LCD display |
|
|
|
|
| Weight:0.8Kg |
|
power waste:12W |
|
|
|
|
| Power supply:Lithium ion battery DC7.4V/continuous working time 6 hours |
|
Particle size channel:0.3、0.5、5µm |
|
|
|
|
| Display unit:Particle/2.83 liters, particle/cubic meter optional |
|
Sampling traffic:2.83L/min |
|
|
|
|
| Sampling period1-40min |
|
Alarm level: ISO5、ISO6、ISO7、ISO8 |
|
|
|
|
| printExternal rechargeable thermal printer |
|
Data storage:256 sets of samples |
|
|
|
|
| Self purification time:≤10min |
|
Standard accessories:Power adapter, printer (including battery and charger), constant speed sampling head, sampling tube, self-cleaning filter, manual, certificate of conformity, factory inspection report, warranty card |