1.1 Vacuum and helium dual optical paths
1.2 Diversified product tray design
1.3 X-Ray detector
1.4 Multiple specifications of test light spots
1.5 Small details, great wisdom
Detector:
●Type: X-123 detector (using original imported high-performance electrically cooled semiconductor detector)
●Be window thickness: 0.5mil
● Crystal area: 25mm2
● Best resolution: 125eV
● Signal processing system DP5
X-ray tube:
Voltage: 0~50kV
Maximum current: 2.0mA
Maximum power: 50W
Target material: Mo
● Be window thickness: 0.5mm
● Service life: greater than 20000 hours
high-voltage power supply:
Output voltage: 0~50kV
● Filament current: 0~2mA
Maximum power: 50W
Ripple factor: 0.1% (p-p value)
8-hour stability: 0.05%
camera:
● Micro focal length
● Drive free
5 million pixels
Collimator, filter:
Quick disassembly of collimator and filter system
● Multiple material collimators
Spot size: Φ 0.7mm, Φ 1.2mm, Φ 3.0mm, Φ 5.0mm
Multiple combinations of filters and collimators, automatically switched by software
Vacuum pump:
● Double stage rotary vane vacuum pump
Power: 400w
● Suction rate: 8m3/h
Other accessories:
Imported high-performance switching power supply
Imported low-noise, high air volume fan
● Dimensions: 470 mm x 560mm x 1030 mm (length x width x height)
Sample chamber size: Φ 188mm × 88mm (diameter x height)
● Instrument weight: 95kg
Power supply: AC220V/50Hz
Maximum power: 1000W
Working temperature: 15-30 ℃
Relative humidity: ≤ 85%, no condensation
Alloy analysis
Can accurately detect elements such as Al, Si, P, S, Cu, Fe, Zn, Ni, Cr, Pb, Mn, Ag in alloys such as copper, steel, and tin
For light matrix alloys, it has excellent detection capability
Zero halogen detection
Vacuum extraction eliminates the interference of Ar in the air, achieving unprecedented stability and accuracy in chlorine measurement!
Toy safety
● To cope with the detection of Pb, Cr, Ba, Hg, Cd, Sb, Se, As, and 8 major elements in EN71-3 and ASTM963-11 instructions, as well as the content detection of Pb, Cr, Ba, Hg, Cd, Sb, Se, As, Co, Cu, Zn, Mn, Sr, Sn, and Ni elements in the new toy instructions.
Coating thickness measurement
Various substrates plated with Au, Ag, Cu, Ni, Sn, Zn, Cr, etc
Response to RoHS Directive
Test results will be available in 60 seconds
World class stability and accuracy;