Dual channel pulse power meter YS2499B
YS2499B dual channel pulse power meter is a newly upgraded multifunctional dual channel digital pulse peak power meter, which has the ability to measu
Product details

YS2499B Dual Channel Pulse Power Meter


1overview

YS2499B Dual Channel Pulse Power Meter(Dual channel waveform microwave power meterIt is a newly upgraded multifunctional dual channel digital pulse peak power meterCapable of measuring pulse signals and continuous wave signals.YS2499B Dual Channel Pulse Power Meter(Dual channel waveform microwave power meterBuilt in dual channel system12bit80MSampling rate: high-speed samplingADCAnd cooperateFPGAThe new generation pulse high-speed signal processing method can accurately reproduce the detection envelope of the measured signal without distortion;YS2499B Dual Channel Pulse Power Meter(Dual channel waveform microwave power meterofThe measurement interface display adopts4.3The inch full-color LCD display screen, as the display interface, can accurately restore the waveform of the measured signal.YS2499B Dual Channel Pulse Power Meter(Dual channel waveform microwave power meterconfigurationLAN(This feature is not enabled as standard)USBRS232Programmable interface, diversified digital communication interface for users to remotely control. The back panel of YS2499B dual channel pulse power meter (dual channel waveform microwave power meter) is also equipped with two additionalBNCThe interface is used forA.BThe detection outputs of two channels can be provided to an oscilloscope for real-time monitoring of the detection envelope of the measured signal.

YS2499B Dual Channel Pulse Power Meter(Dual channel waveform microwave power meterIt can be achieved through external devicesTwo channels measure the incident power separatelyAnd reflected powerThus, corresponding standing wave ratio and reflection system can be obtained, and continuous wave microwave power can also be measured.YS2499B pulse power meter(Dual channel waveform microwave power meteralsoUsed for standing wave ratio monitoring of various microwave high-power systems.YS2499B pulse power meter(Dual channel waveform microwave power metercorrectThe upgraded product of YS (M) 2498 dual channel pulse RF power meter.Users can choose specific directional couplers to expand the power measurement rangeIt can be measured up to 100MW

twocharacteristic

1. Can test the peak power of pulse signals.

2. The power test units dBm and W can be converted to each other.

3. Capable of manual zeroing and power correction.

4. Frequency range: 50MHz-4GHz.(Frequency range can be customized according to requirements, up to 12.4GHz)

5. Power range: 1uW-100mW.(Customizable power range according to requirements)

6. Accuracy: ≤ ±0.5dB

7. Pulse: Width>1 μ s(Customizable according to requirements), repetition frequency50Hz-20kHz.

8. Input interface: N-type.

9. Detection function:

1) It has a detection output port with a Q9 head. When connected to an oscilloscope, this detection port can display the detection waveform of pulse RF signals on the oscilloscope.

2) When the input power is ≥ 100mW, the voltage amplitude of the detection output port is ≥ 0.8V.

3) The detection circuit can fine tune the rising and falling edges, but this function is not reflected outside the device.

10. Work environment:- 10 ℃ -50 ℃.

11. Power supply: mains power.

IIMain technical parameters

Technical indicators

Typical technical parameters

1

Frequency range

50MHz~4000MHz

2

Power measurement range

-30dBm~20dBm1μw ~ 100mW

3

Tested signal * narrow pulse width

1uSCustomizable according to requirements

4

Modulation period range of the tested signal

0.4uS~CW

5

measurement error

≤±0.5dB

6

communication interface

LAN (this feature is not enabled as standard)USB (optional)RS232

7

User selected interface

A.BChannel detection monitoring output

8

display

480*272Full Color-4.3cun

9

power supply

AC110V-240V 50Hzperhaps60Hz

10

Dimensions

220mm×120mm×280mm(Length x Width x Depth)

Note:Ys2499BAs a universal frequency band measuring instrument, the measurement frequency band can be customized according to the actual needs of users,Expand the power range,And other customized functional services can be provided according to actual situations.

IIIAdjust the control mechanism

3.1 LCDdisplay

Display waveform, power, and related parameters

3.2 Power sockets and switches

When the power is plugged in and the switch is turned on, the instrument is powered on

3.3~3.4 Keyboard

Instrument workers include11A button, which are4A function key4One directional key, logarithmic number/Watt key, waveform/Data and menu keys

3.5 AChannel input port

Connect through this input portAChannel input power (generally considered as incident power)

3.6 BChannel input port

Connect through this input portBChannel input power (generally considered as reflected power)

3.7 LANopening

3.8 RS232interface

3.9 Incident channel detection monitoring alarm interface

outputCHAThe envelope signal of signal detection can be used for oscilloscope monitoring

3.10Reflection channel detection monitoring alarm interface

outputCHBThe envelope signal of signal detection can be used for oscilloscope monitoring


4Introduction to Screen Display Parameters

4.1 T

cursor1With cursor2Horizontal spacing (i.e. the time difference between two pointers)

4.2 Mpos

Location of the main time base (default to the center position, this feature is not yet upgraded)

4.3 Mark1

cursor1correspondingCHAandCHBThe power of

4.4 Mark2

cursor2correspondingCHAandCHBThe power of

4.5 Avg(A)AvgB

ABthoroughfareMark1giveMark2The average power between them

4.6 Peak

ABPeak power of the channel

4.7 Menu Area

5Usage

5.1 Precautions

a. Before using this instrument, one should first read the instrument manual and become familiar with the operating methods;

b. This instrument usually works normally after a few minutes of startup. For precision testing, the preheating time is 15 minutes.

5.2 Button Introduction

Ys2499BOn the front panel, there are11 buttons, including 4 function keys, menu key, zero adjustment key, calibration key, and 4 directional keys. There is a power button at the power input interface on the back panel that controls the power on/off.

a. Function keys

Ys2499BshareFour function keys are used to select submenus under each menu page.

b. Menu key

This key is mainly used for switching between menu pages.

c.logarithm/Watt switching

This key is used for logarithms/Watt switching display

d.wave formdata

This key is used for displaying waveforms on the instrument/Switching between data

e. Direction keys

Ys2499B has four directions: up, down, left, and right. The up and down keys are mainly used for numerical addition and subtraction, while the left and right keys are mainly used for cursor movement under submenus.

5.3 Menu Introduction

Ys2499BJointly set up5 menus, namely Trigger Channel A、Channel B、Marker、Alarm&232、Lan、ZERO/CAL, They switch through menu keys, pressing the corresponding function key on the right side of each submenu, and then using the up, down, left, and right arrow keys to operate the submenu or set the corresponding parameters.

Trigger

Trig Mode

It has three triggering modes, namely automatic, normal, and single trigger

Trig Source

Trigger sources: CHA/CHB/NEXT,AT1912B has external triggering function, but does not reserve panel interface

Trig Level

Trigger level: Trigger level range -30dBm~+20dBm

Frequency

Set measurement frequency to compensate for amplitude frequency response

Channel A

Channel

thoroughfareAThe waveform displayON/OFF

Ver Center

The waveform displays the reference level in the center

Ver Scale

Vertical resolution of waveform display

Offset

Set the offset for measurement compensation

Channel B

Channel

thoroughfareBThe waveform displayON/OFF

Ver Center

The waveform displays the reference level in the center

Ver Scale

Vertical resolution of waveform display

Offset

Set the offset for measurement compensation

Marker

Marker1

cursor1, used for measuring markers

Marker2

cursor2, used for measuring markers

Time Base

Set horizontal resolution

Hor Center

Set the starting reference line for waveform measurement display

Alarm&232

SWR Alarm

Set the standing wave alarm value toCHAFor incidence,CHBFor reflection (function not enabled)

Ref Alarm

set upCHB*High measurement power alarm power (functional, no reserved interface on the panel)

Baudrate

set upRS232/USBThe communication baud rate

Uart Mode

RS232/USBcommunicationON/OFF

Lan

IP Addres

Network port parameter settings

Subnet Mask

Gateway

Enter

Confirm and save parameter settings

ZERO/CAL

ZERO A

CHAZero adjustment

Calibration A

CHACalibration, during calibration, the reference source automatically opens, and after calibration is completed, the reference source automatically closes

ZERO B

CHBZero adjustment

Calibration B

CHBCalibration, during calibration, the reference source automatically opens, and after calibration is completed, the reference source automatically closes

VICommunication Interface Protocol

USBInterface: (This)USBFor VirtualUSBInterface, similar toRS232COMMUNICATION FORMAT, N.8.1

function

instruction

*range

1

Frequency setting

ASA_[FREQ](**.**GHz)

09.30~09.50

2

offset setting

ASA_[OFFSET](±**.**,±**.** dB)

-99.99~+99.99

3

Trigger level

ASA_[TRIG](±**.** dBm)

-30.00~+20.00

4

Trigger Mode

ASA_[MODE](*)

0,1,2auto,normal,single

5

trigger channel

ASA_[CH_X](*)

0,1,2CHA,CHB,EXT

6

Scan time

ASA_[TIME](**)

2~17

7

Read measurement values

ASA_[RETURN](*)

Any: Default“0

Network communication protocol (following)ModBusProtocol format)

Set instruction format

function

instruction

_meaning

1

Frequency setting

41 53 41 10 A1 02 _ _CRCH CRCL

2One byte,50MHz~4G

Range:

0x00 0x05~0x01 0x09

2

offset setting

41 53 41 5F A2 04 _ _ _ _ CRCH CRCL

front2One isCHA

after2One isCHB

defined as2Bytunsigned number

3

Trigger level

41 53 41 5F A3 02 _ _ CRCH CRCL

Defined as unsigned2individualBytunsigned number

4

Trigger Mode

41 53 41 5F A4 01 _ CRCH CRCL

0,1,2auto,normal,single

5

trigger channel

41 53 41 5F A5 01 _ CRCH CRCL

0,1,2CHA,CHB,EXT

6

Scan time

41 53 41 5F A6 01 _ CRCH CRCL

2~17

7

Read measurement values

41 53 41 5F A701 _ CRCH CRCL

arbitrary

Return instruction format:

The system automatically measures parameters:CHA,CHBofPeakprice

41 53 41 5F B1 04 _ _ _ _ CRCH CRCL

explain:

Data header: 41 53 41 5F

Function code:B1

Data length:04

data

16Bit verification code:CRCH CRCL

The format for setting and returning instructions is the same.