CLJ-3016L Stainless Steel Slant AC/DC Dual purpose Laser Dust Particle Counter
This instrument is designed in accordance with the requirements of the new GMP (2010 version). The overall material of this product is made of 316L stainless steel shell. The laser sensor adopts a full semiconductor laser and a semiconductor photodiode receiver to ensure the stability of the light source and the accuracy of the signal receiver. Greatly reducing the stray light inside the scattering cavity and improving the signal-to-noise ratio of the sensor. Adopting LCD large screen display and PVC film switch control, the display data is clear and the operation is convenient. This instrument can automatically determine various purification levels based on the standards selected by the customer. This product is widely used in medical, pharmaceutical, electronic, chemical, optical, food, cosmetics, biological products, aerospace, and various testing institutions for the detection of dust particles
Main technical parameters:
light source |
Semiconductor laser with a lifespan greater than 10 years |
sample flow |
2.83L/min (0.1ft3) |
Display mode |
122 × 90mm high-resolution LCD screen display (LCD) with Chinese and English interface selection, real-time display, previous cycle display, real-time concentration display, can display time, date, measurement value, temperature and humidity, room number, sampling point, sampling frequency, battery level, status and other parameters, 95% UCL calculation, can directly display particle concentration (particles/cubic meter) |
Rechargeable Battery |
Lithium ion battery, 16.8 V, 2600mAh |
power supply |
AC power adapter, AC: 100V~245V, 50/60 Hz to DC: 16.8V, 1A; |
working hours |
8 hours (remaining battery indicator) |
count mode |
Cumulative value, difference value, concentration value |
testing mode |
Single, repetitive, continuous, computational, remote |
Units |
The unit can be converted into m3, ft3 |
Particle size channel |
0.3 、 0.5、 1.0、 3.0、 5.0、 10μm, Simultaneously count six particle sizes |
sampling period |
1-10 (min) Delay count: 0-99 (S) Self purification time: ≤ 10 (min) |
work environment |
temperature:0~40℃(50~104℉)Relative temperature:20~90%RH, No condensation Atmospheric pressure: 86-106KPa |
Temperature and humidity: |
Choose (1) Temperature: 0-50 ℃± 1 ℃ (2) Humidity: 0-100% RH ± 5% |
UCL settings |
Sampling points (A): Set from 2 to 9 points. Sampling frequency per point: (L) 3-5 times Set measurement position: 0-999 |
UCL report |
Compliant with ISO14644-1, GB50073-2001, GMP (2010) |
Data Storage |
Can store 1500 sets of data(Circular buffer)(including particle size, data, environmental data, year, month, day, time, sampling volume, data location)Data will not be lost after power failure. |
communication interface |
RS232, 9600 baud rate |
Alarm settings |
The instrument is equipped with a level alarm function, which can sound an alarm when the cleanroom exceeds levels 100, 1000, 10000, 100000, 300000, and 1000000, Measurement of static and dynamic standards at four levels A/B/C/D that comply with the new GMP standards/switching measurement of four standards |
zero count |
Meets the requirements of JJG-547-88 Dust Particle Counter Verification Regulations JIS B9921:1997, with less than one count every five minutes |
Basis Set Superposition Error |
At 5%, 2000000 particles per cubic foot |
Printing function |
Built-in printer |
External dimensions |
260×130×220 |
weight |
2.6kg |
calibration |
Traceable to the National Institute of Standards and Technology (NIST) in the United States, our company has passed the national metrology standard building assessment and can be traced back to the Shanghai Institute of Metrology and Measurement Technology. We can also calibrate ourselves or third-party national metrology institutions |
Standard configuration |
Suitcase/charger, power sampling head, zero filter, USB to RS 232 module, RS 232 connection cable, computer communication software CD, built-in printer, professional triangular sampling rack, |
Optional parts |
temperature and humidity sensor |