CJJ-931 Double Digital Constant Temperature Quadruple Magnetic Stirrer
Double digital constant temperature four link magnetic stirrer 1. Power supply voltage: 220V 10% 50HZ2. Motor power: single motor 25W3. Hot wire power
Product details

CJJ-931 Series Digital Constant Temperature Magnetic Stirrer User Manual

1、 Dual digital constant temperature quadruple magnetic stirrerbrief introduction
This instrument is widely used in various universities, environmental protection, scientific research, health, epidemic prevention, petroleum, metallurgy, chemical, medical and other units. According to the requirements of the majority of users, it is designed with multi connected heating, stirring and constant temperature control such as two connected, four connected, and six connected. Enable users to obtain corresponding effects and data for multiple samples simultaneously. This instrument has good performance, low noise, and stirring, making it an ideal tool for laboratory technicians.

IIDual digital constant temperature quadruple magnetic stirrerperformance

1. Power supply voltage: 220V ± 10% 50Hz
2. Motor power: 25W per motor
3. Hot wire power consumption: a single hot wire can consume up to 300W
4. Magnetic steel speed: maximum 2000 revolutions per minute
5. Digital temperature control: room temperature~100 ℃± 1.0 ℃
6. Digital display speed: starting~2000 revolutions per minute, accuracy ± 5 times per minute
7. Can be mixed synchronously or separately, mixing capacity: 6 × 1000ml

3、 Usage method
When using this instrument, please first check if all the accessories accompanying the machine are complete. Then, install the fixtures in order and place the beaker that needs to be stirred in the center of the plating plate. Add the solution and place the stirrer in the beaker solution. Then, plug in the power plug of the instrument first, turn on the power and turn on the power speed control switch. The display will change from 99.9 to the actual temperature inside the instrument container. Press the "SET" button to make the digital tube on the right decimal point of the display flash. At this time, set the desired preset temperature. Press ▲ to increase the number, and pressReduced. Press the "SET" button again, and the data will be displayed as the actual temperature inside the instrument container. If the temperature of the instrument's contents does not match the displayed temperature, make the following adjustments: long press the "SET" button, when the display shows "E 1.5", release it (E 1.5 is the advance control, which has been adjusted before leaving the factory, do not adjust it by yourself), then press the "SET" button again, the number display will appear (there are no English letters at this time), press ▲ to increase the temperature, pressReduce the temperature (for example, if the actual temperature inside the instrument container is greater than the displayed temperature, press the ▲ key to adjust the deviation between the two), and continue calibration. Press the "SET" button again, and it will display "P 2.0" (P 2.0 is a proportional band that has been adjusted before leaving the factory, please do not adjust it yourself). Press' SET 'again to display the temperature inside the instrument container. The 'OUT' light indicates that the preset temperature is higher than the temperature inside the instrument and is in a heating state. The speed regulation is gradually adjusted from low speed to high speed, and it is not allowed to start the high speed gear directly to avoid the agitator not being synchronized and causing jumping. Temperature control: Insert the probe into the socket and adjust the temperature control knob to the desired temperature. When not working, cut off the power supply. To ensure safety, please connect the ground wire when using. The instrument should be kept clean and dry, and it is strictly prohibited for solution to enter the machine to avoid damaging the parts and prevent severe vibration.

4precautions
1. If you find that the stirrer is jumping or not stirring during mixing, please cut off the power and check if the bottom of the beaker is flat. The position is

No, it's not correct. Please also test if the current voltage is between 220V ± 10V, otherwise the above situation will occur.
2. The heating time should generally not be too long, intermittent use can extend the service life.
3. Medium speed operation can work continuously for 8 hours, while high-speed operation can work continuously for 4 hours.

5List of complete machines

1. 1 complete machine 2, 6 stirrers 3, 1 power cord 4, 6 chrome plated uprights

5. 6 chrome plated cross joints, 6 rubber clamps, 7 rubber wood support screws, 12 rubber wood fastening screws

8. 1 manual 9, 1 warranty card