BINDER MKF56 High and Low Temperature Alternating Climate Chamber
1、 Product Introduction:
1. Temperature range: -40 ℃ to 180 ℃
2. Humidity range: 10% relative humidity to 98% relative humidity
3. Four zero voltage switch contacts that can be activated through the MCS controller
4 APT.lineTMPreheating chamber technology
5. Programmable condensation protection for samples
6. Heated observation window with LED internal lighting
7. Using capacitive humidity sensors and steam humidification for humidity regulation
8. BINDER communication software APT-COM 3 Basic version
9. Fault diagnosis system using optical and sound alarms
10. Intuitive touch screen controller using interval programming and real-time programming
11. Internal data logger and open format measurement values can be read through USB
12. Equipment self-test is used for comprehensive status analysis
13. Access hole with silicone plug 50mm, left
14. 4 casters (2 with brakes)
15. Computer interface: Ethernet
16. Adjustable slope function function function
17. Integrated continuous automatic recorder
18. Real time clock
19. Door heating
20. 1 socket made of stainless steel
21. Hose protection joint, including water supply and drainage pipes with a height not exceeding 1m
22. Inner chamber made of stainless steel
23. PT100 temperature sensor
24. Heat insulation through polyurethane foam
25. Fluorine free refrigerant R-404A
26. Cooling through a compression cooler
27. Finned evaporator
2、 Technical parameters:
| Product Model | MKF056-230V1 | MKF056-240V1 |
| Choose a model | standard | standard |
| Item code | 9020-0317 | 9020-0396 |
| measurement | ||
| Internal volume [L] | 60 | 60 |
| Net weight of equipment (unloaded) [kg] | 175 | 280 |
| Maximum total load [kg] | 60 | 60 |
| Maximum load per partition [kg] | 15 | 15 |
| The dimensions of the casing do not include additional components and connectors | ||
| Net weight width [mm] | 720 | 720 |
| Net weight at height [mm] | 1445 | 1445 |
| Depth net weight [mm] | 780 | 780 |
| Wall distance from rear [mm] | 300 | 300 |
| Wall to side distance [mm] | 200 | 200 |
| Observation window width [mm] | 288 | 288 |
| Observation window height [mm] | 255 | 255 |
| internal dimensions | ||
| Width [mm] | 400 | 400 |
| Height [mm] | 420 | 420 |
| Depth [mm] | 348 | 348 |
| The number of doors | ||
| inner door | 1 | 1 |
| outer gate | 1 | 1 |
| Data related to the environment | ||
| Sound pressure level [dB (A)] | 59 | 59 |
| firmware | ||
| Number of partitions (hours/maximum) | 1/4 | 1/4 |
| Temperature performance data | ||
| Temperature range [℃] | -40…180 | -40…180 |
| The temperature deviation depends on the set value [± K] | 0.5…1.5 | 0.5…1.5 |
| The temperature fluctuation depends on the set value [± K] | 0.1…0.5 | 0.1…0.5 |
| The average heating rate is based on IEC 60068-3-5 [K/min] | 5 | 5 |
| Cooling time from 180 ℃ to -40 ℃ [min] | 90 | 90 |
| The average cooling rate is based on IEC 60068-3-5 [K/min] | 5 | 5 |
| The maximum thermal compensation is at 25 ℃ [W] | 800 | 800 |
| The maximum thermal compensation is at 25 ℃ and 90% relative humidity [W] | 250 | 250 |
| Climate performance data | ||
| Temperature range [℃] | 10…95 | 10…95 |
| The temperature fluctuation depends on the set value [± K] | 0.1…0.5 | 0.1…0.5 |
| Humidity range [% relative humidity] | 10…98 | 10…98 |
| The humidity fluctuation depends on the set value | ≤ 2.5 ± relative humidity | ≤ 2.5 ± relative humidity |
| Condensation temperature range of condensate water [℃] | 5…94 | 5…94 |
| Electric data | ||
| Rated voltage [V] | 200…230 | 200…240 |
| Power frequency [Hz] | 50 | 60 |
| Rated power [kW] | 2.8 | 2.8 |
| Equipment safety device [A] | 16 | 16 |
| Phase (rated voltage) | 1~ | 1~ |
1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.