Autoclave for cement stability experiment
The autoclave used for cement stability testing meets the new national standards and is a specialized instrument used by cement production units, buil
Product details

Product Usage:

Autoclave for cement stability experimentApplicable to the national standard GB/T 750, it is a specialized instrument used by cement production units, building materials research institutes, colleges and universities, and construction quality inspection departments to determine the adverse effects of MgO (magnesium oxide) in cement on cement stability.

The performance and testing of cement indicate that the content of MgO in cement clinker is mainly brought in by limestone raw materials, and most of the MgO in clinker is in the form of periclase after firing. Its structure is dense (commonly known as dead burning state) and wrapped in the middle of clinker minerals, with a very slow reaction rate with water. It is generally believed that after 10-20 years or more, hydration continues to occur, resulting in the formation of Mg (OH)2Increasing the solid phase volume by 2.48 times, local volume expansion generates significant destructive stress inside the already hardened cement stone. Boiling the test cake at 100 ℃ for 3 hours cannot cause a large amount of hydration of MgO, so the boiling method cannot be used to test the cement stability caused by magnesite. High temperature and high pressure conditions can accelerate the hydration of magnesite in clinker, so the poor stability caused by MgO content in clinker needs to be tested by steaming method.

Technical parameters:

1. Pot volume: 8.5L

2. Inner diameter of kettle body: Φ159mm

3. Work pressure: 2.5MPa (adjustable manually)

4. Standard boost time (0-2 ± 0.05MPa); Room temperature~215.7 ± 1.3 ℃): ≈ 45min

5. Opening pressure of safety valve: 3.0MPa

6. The fluctuation range of automatic pressure control: ≯ ±0.05Mpa

7. Pressure holding time setting range: 1~99h

8. Rated voltage: 220V

9. Rated power: 1200+600W

10. Dimensions: 1000 × 660 × 1320mm

11. Weight: Approximately 75kg

Instrument installation:

1Autoclave for cement stability experimentIt should be installed in a relatively secluded place. To ensure safety and ease of use and maintenance, the instrument should face away from the wall and be no less than 500mm away from the wall.

The instrument should have a reliable grounding wire.