·The air cleanliness in the operating room meets the relevant requirements of ISO 14644.1 level 3
· Mini folding non partition ULPA super efficient filter technology reduces energy consumption and improves laminar flow uniformity, providing better product/sample protection and preventing cross infection
· The unique horizontal uniform laminar flow design continuously provides clean air during operation, thereby preventing pollution particles from depositing in blind spots
· The air passes through high-quality ULPA ultra efficient filters with a 99.999% interception efficiency for particles larger than 0.12 microns. After double filtration, it is discharged to ensure the safety protection of products/samples, laboratory staff, and laboratory environment
·SentinelTMThe microprocessor controller has a soft touch control panel surface for easy cleaning and is equipped with an LCD display screen
·The back wall of the operating room is equipped with Magnehelic®Differential pressure gauge, used to monitor the pressure conditions in the operating room during the use of biosafety cabinets
·Neoprene with world leading technologyTMSynthetic gloves have better flame resistance, abrasion resistance, cleanliness, and chemical corrosion resistance than ordinary latex gloves
·The transfer window installed on the side of the main operating room has a double door interlock function. The items required for the experiment enter the safe cabinet operating room through the transfer window, ensuring that the experimental operation process does not pollute the laboratory environment
·The safety cabinet adopts an ergonomic inclined design, while reducing the reflectivity of the glass surface and improving the operational comfort of the staff
·Stainless steel countertop is durable, corrosion-resistant, and easy to clean
·The built-in mild white, electromagnetic ballast flicker free fluorescent lamp provides the best lighting for the entire operating room area, reducing visual fatigue for operators. The UV resistant tempered glass window and the built-in stainless steel glove operating port are designed with 100% zero leakage
·Esco Patent IsosideTMAntibacterial coating to prevent bacterial and microbial contamination on the surface of the cabinet
·Adopting a unique Dynamic Chamber
TMThe design provides additional safety protection for negative pressure layers outside all contaminated areas