AL-AC130 Ammonia Nitrogen Automatic Monitoring Instrument
This method is suitable for wastewater with ammonia nitrogen in the range of (0-500) mg/L. Excessive calcium and magnesium ions, residual chlorine, or turbidity may interfere with the measurement.
A pollution source installed in a specific location, applied in the field of water quality monitoring, mainly used by sewage treatment plants.
Main performance parameters
Main parameters and technical indicators
Measurement range (0-500) mg/L (expandable)
Quantitative lower limit ≤ 0.15mg/L (indication error ± 30%)
Repeatability ≤ 2%
Low concentration drift ≤ 0.02 mg/L
High concentration drift ≤ 1.0%
The minimum measurement cycle is 20 minutes, and according to the on-site environment, the color development time can be modified arbitrarily within 5-120 minutes
Sampling cycle time interval (adjustable from 10 to 9999 minutes) and whole point measurement mode
Calibration cycle 1-99 days, adjustable at any interval and time
Maintenance cycle>720h/time
Output RS-232, RS-485, 4-20mA (optional)
For indoor environments with adjustable temperature requirements, it is recommended to have a temperature range of+5-28 ℃; Humidity ≤ 90% (non condensing)
Power supply AC220 ± 10% V, 50 ± 10% Hz, 5A
Indication error:
± 8.0% at a standard concentration of 2.0 mg/L
± 5.0% when the concentration of the standard solution is 5.0 mg/L
± 3.0% at a standard concentration of 8.0 mg/L
memory effect
± 0.3 mg/L when the concentration of the standard solution is 2.0 mg/L
± 0.2 mg/L when the concentration of the standard solution is 8.0 mg/L
Actual water sample comparison test
Water sample concentration<2.0 mg/L ≤ 0.2 mg/L
Water sample concentration ≥ 2.0 mg/L ≤ 10.0%