Online pH analyzer
model: P53A4A1N
Specifications:
Range: PH-2.00-14.00, customizable
Temperature:- 20C--200C
Accuracy: 0.1% of the measuring range
Power supply: 220VDC, 50HZ
Output: 4-20mA
Electrode options: 6028PO or PD1R1 or PC1R1A
6028PO differential pH electrode/2028RO differential ORP electrode
Range: 0-14PH/-2100-2100mV
Resolution: better than 0.005pH/0.5 mV
Temperature:- 5?C—95? C
Pressure resistance: 100PSI
Liquid bonding material: LCP body, PVDF reference junction, Viton O-ring, glass electrode,
Titanium grounding shaft
*Large transmission distance: 914 meters
The GLI differential electrode has very high accuracy and excellent reliability. The electronic circuit is completely sealed, with good moisture-proof effect. The electrode shell material is liquid crystal polymer, with good chemical stability and corrosion resistance. The dual stage reference electrode can resist toxicity and ensure reliable testing. The original factory provides a one-year quality guarantee. Applied to general sewage, its service life can be extended by 12-15 years.