Unlike traditional sampling techniques such as sedimentation, PurCon samplers can provide online analyzers withContinuous flowing, particle free, suspended solids, and microbial leachateTherefore, it can effectively avoid pipeline blockage and pollution, ensuring the reliable operation of the entire system.
In addition, due to the sampling method complying with the German DIN38405/38406 laboratory standard testing method, the online test results have good consistency compared to the laboratory test results.
Due to the use of special sampling techniques,Only require a short processing timeThis ensures that the test results provided by the analyzer have strong real-time representativeness, reflecting almost the actual concentration in the current medium, and is more suitable for real-time monitoring of emission systems.
Compared with general sampling systems, such as ultra pure filtration equipment, PurCon has very low installation and investment costs. In addition, PurCon hasSelf cleaning function,It can greatly extend the maintenance cycle, such as once every 9 months.
PurCon is not only designed to be used in general settings such as biological purification processes, but also in wastewater treatment plants with complex sedimentation characteristics using activated sludge treatment processes. Field practice has proven that PurCon is an effective and reliable sampling system.
Main technical parameters
Permeation fluid flow rate: continuous, maximum 3.6 L/hour, adjustable in four ranges, free of particles and bacteria
Submersible pump: 400 1500 L/hour
Connector: Sample inlet
3/4 "hose connector with inverted hook
Drainage outlet 1.97 "PVC pipe joint
Cleaning solution outlet
3/4 "hose connector with inverted hook
Penetrating fluid outlet 0.06 "threaded joint
Voltage: 230VAC/115VAC
Power consumption: 150VA
Material, weight: 316 stainless steel, 36 kg
Aperture: 0.4 μ m in accordance with DIN 38405+38 406
Transmission method: continuous
Lag time:< 5="" min ="" t100="">< 7="" min="">
Maintenance requirement: less than 1 hour every six months
Film lifespan:> 8 years
Power consumption PurCon: 150 W