The inspection and calibration of the position of the light source emitter (filament) in the microscope. This article is organized and provided by an
Product details

Inspection and calibration of the position of the light source emitter (filament) within the microscope
This article is organized and provided by an inverted microscope

The purpose is to align the image end of the luminous body with the field of view of the objective lens, ensuring sufficient and uniform illumination of the microscope's field of view from the perspective of the light source. This is a prerequisite for adjusting the Kuller illumination system. Basic tools required: The telescope is equipped with a microscope when purchasing it.
 
Remove the frosted glass sleeve from the lamp compartment and reinstall the lamp chamber onto the microscope;
select10X objective lens, turn on the light source program to find the sample and focus it clearly, then switch to another one40Focus the sample clearly with the objective lens(40The objective lens can clearly see the entire picture of the filament;
Open both the aperture stop and field stop of the spotlight to their maximum size;
Pull out one of the eyepieces, replace it with a centering telescope, grasp the white part, and stretch the black eyepiece with the other hand to see the filament image in the field of view;
If the position of the filament is not suitable, adjust the "-" hole to adjust the filament image horizontally, adjust the "-" hole to adjust the filament image vertically, until the filament image is adjusted to just fill the circular image of the objective lens aperture;

After the adjustment is complete, insert the frosted glass sleeve back into its original position, remove the centering telescope, and replace the telescope for the next adjustment. The adjustment of the lighting source chamber outside the microscope and the calibration of the position of the light source emitter inside the microscope as described above only need to be carried out during the initial installation, debugging, and replacement of the light bulb of the microscope. It is not allowed to adjust or move the microscope randomly during normal use. In case of disorder, you can follow the above steps to restore it to its original state.