application area
Widely used in industries such as semiconductors, photovoltaics, metallurgy, and fluorine chemical manufacturing, including wastewater and exhaust gas treatment, emission monitoring, and various ion monitoring.
Instrument features
96 * 96 microcomputer designed ion concentration controller
◆ Outdoor waterproof design suitable for wall mounted, panel mounted, and circular pipe installation
◆ Provide electrode automatic cleaning device contacts, program setting output
Large image LCM display screen with operation status and special symbol display
◆ Equipped with backlight sensor, automatic and manual backlight setting function
◆ Provide solution grounding function, which can eliminate the interference of charged samples
◆ Provide digital filtering function to filter out interference values, making water quality testing more stable
◆ Equipped with automatic calibration; Manual/automatic temperature compensation function
◆ Password lock and combination key enter the setting, calibration mode design, and increase security protection
Equipped with RS-485 standard MODBUS communication protocol, it can be directly connected to digital systems (MB-200-ION-1RS serial only)
Two sets of programmable outputs (MB-200-ION-1 serial only)
Technical Parameter
Measurement range: 0.00~20000 (ppm, mg/l)/-2000.0~2000.0mV/-30.0~130.0 ℃
Resolution: 1/0.1/0.01 (ppm, mg/l)/0.1mV/0.1 ℃
Instrument accuracy ± 0.06% ± 1 Digit/± 0.02 ℃± 1 Digit with temperature error correction function
Reproducibility ± 0.05% ± 10% of the reading value (depending on the ion selective electrode and sample)
◆ Digital filter adjustable from 0 to 60 seconds/factory set value of 30
◆ Temperature compensation automatic temperature compensation PT1000/NTC30K, automatic identification; Manual temperature compensation: button setting
Correction mode: two-point or three-point correction, offset value correction (thirteen sections available for selection)
Input impedance>1012 Ω
Relay output with two sets of ON/OFF contact outputs, high and low points, and hysteresis can be set by any program, 240VAC/0.5A
◆ Analog output 2 channels of 4~20mA analog signal output, users can set the corresponding output value, load 500 Ω
Basic parameters
◆ AC power supply 100V~240VAC ± 10%, 50/60Hz
◆ DC power supply 20-40VDC
Working temperature: 0~50 ℃
Instrument size 96mm × 96mm × 138mm (H × W × D)
◆ Protection level meets IP65 (NEMA 4)
Installation methods include wall installation, panel installation, and circular pipe installation
Excavation size 91mm × 91mm (H × W)
Weight 0.4 Kg
Electrodes and accessories