Instrument Introduction:
HG-SHP seriesIntelligent biochemical incubators are widely used for the cultivation and preservation of bacteria, fungi, microorganisms, and tissue cells, as well as water quality analysis and BOD testing, suitable for breeding experiments and plant cultivation. It is laboratory equipment for research institutions, colleges, production units or departments in the fields of biology, genetic engineering, medicine, health and epidemic prevention, environmental protection, agriculture, forestry and animal husbandry.
Technical parameters:
1. Studio capacity: 150 liters, 250 liters
2. Temperature range: 5-50 ℃
3. Temperature uniformity: ≤± 1 ℃
4. Temperature fluctuation: ≤± 1 ℃
5. Stainless steel liner, plastic sprayed steel plate shell, polyurethane foam insulation
6. Equipped with a large viewing angle observation window and lighting fixtures
7. Microcomputer control, high digital display accuracy (optional LED display screen).
8. Unique stainless steel circulation air duct, forcing air circulation and uniform temperature
model:HG-SHP150、HG-SHP250
instructions:
1. Connect the power supply, turn on the power switch, and the whole machine will be powered on. The power indicator light inside the switch will light up.
2. Please refer to the manual of the intelligent digital temperature controller for temperature control settings.
3. If lighting is required, turn on the lighting switch.
4. This incubator has a breakpoint protection function and a delay function of about one and a half minutes. It takes about one and a half minutes to restart the compressor after it stops.
Maintenance and cultivation:
1. The outer shell of the incubator should be reliably grounded.
2. The incubator should be placed in a cool, dry, well ventilated place away from heat sources and direct sunlight. Place it steadily to prevent vibration and noise.
3. To ensure effective heat dissipation of the condenser, the distance between the condenser and the wall should be greater than 100mm. There should be a 50mm gap on the side of the box, and at least 300mm of space at the top of the box.
4. When handling, repairing, and maintaining the incubator, collision, shaking, and vibration should be avoided; The maximum inclination is less than 45 degrees.
5. The instrument suddenly doesn't work. Please check if the fuse tube (behind the box) is burnt out and check the power supply.
6. When cooling this incubator, it is not advisable to make the temperature difference between the incubator and the ambient temperature greater than 25 degrees.