Marsurf UD 120 Combined Contour and Roughness Measuring Instrument from Germany
German Marsurf LD 120 combination contour and roughness measuring instrument
German Mahr Marsurf UD 120 combined contour and roughness measuring instrument
German Mahr Marsurf LD 120 combined contour and roughness measuring instrument
Mahr MarSurf UD 120/LD 120 from Germany combines contour and roughness measurement in one
By adopting Mahr's leading metrology technology, contour measurement and roughness measurement can be solved simultaneously after one measurement. Adhering to this excellent philosophy, Marsurf UD120 and Marsurf LD120 can provide you with high-precision measurements with nanometer level resolution. The patented technologies include: probe replacement device, automatic recognition system based on measurement operation, etc
Combination based contour and roughness measurement
·German Mahr MarSurf UD 120 combination contour and roughness measuring instrument combination
The Marsurf UD 120 combined contour and roughness measuring instrument is a universal basic equipment for evaluating contour and roughness in one measurement, with nanometer level resolution.
·German Mahr MarSurf LD 120 Combination Contour and Roughness Measuring Instrument
The German Mahr MarSurf LD 120 roughness profile measuring instrument has been widely praised in the customer market for many years, representing the leading technology in contour and roughness measurement. Its excellent technical parameters, such as residual error Rz=35 nm and dynamic force control range of 0.5~30mN, ensure that the equipment can achieve the required accuracy, even for the highest precision workpiece measurement requirements.
Supported by universal software and automatic control concepts, the Marsurf UD 120 combined contour and roughness measuring instrument provides a contour and roughness measurement station with Mahr's leading technology. Software control is built on a software platform now known as "MarWin", which achieves modularity and universality, and is easy to operateMake it. The Marsurf UD 120 roughness profile measuring instrument is randomly equipped with the XCR 20 CNC software package, providing a wide range of measurement and evaluation functions, especially for profiles and roughness. The drive shaft and column lifting shaft can be controlled by sitting on the same operating lever or through program control.
·Explanation
-Complete roughness and contour evaluation in one measurement
Less time, faster operation, more reliable results, and more efficient work, all roughness and contour parameters can be obtained with just one measurement. Leave you more time to complete other tasks!
-The most modern conversion technology is more suitable for small tolerance workpieces
Due to the device's ability to achieve high-precision measurements with high resolution over a large measurement range, you can leave smaller allowable errors for your machining tasks and take on new challenges.
-Innovative arm movement concept achieves maximum stability
The mobile measuring arm adopts a bridge structure design, ensuring maximum rigidity, minimum measurement vibration, and flatness; In this way, during use, you can construct bridge structures of any size!
-Magnetic suspension bracket for measuring arm
Magnetic installation design can ensure high-frequency use of the system, collision safety protection, and achieve long service life.
German MahrPrinciple of MarSurf UD 120/LD 120 Measurement Station
·Components of the measuring station
Drive MarSurf UD 120/LD 120
The Marsurf UD 120/LD 120 driver can easily achieve a measurement stroke of 0.1mm to 120mm, with speed control within 0.1 to 2mm/s. Due to the high-precision grating ruler used on the X-axis, automatic positioning can be achieved.
·Working principle of probe system
The working principle of the probe is based on a joystick system with a measuring tip, and the other end of the joystick is a diffraction grid, which can measure the swing angle of the joystick. The light emitted by the light-emitting diode shines on the diffraction grid, producing an interference pattern. The rise and fall of the measuring tip will cause changes in the interference pattern at the receiving end, and the subsequent evaluation circuit can determine the position of the grid, which in turn determines the position of the measuring tip.
A mobile rotary motor is used for Z-direction probe positioning to determine the direction of probe movement and achieve force control.
To meet most common applications, the Marsurf LD120 probe set is equipped with LD A14-10-2 90 ° by default, LD A14-10-500 (LD 120), Several measuring arms, including LD A14-10-2 60 ° (UD 120). The rigid skeleton structure of the measuring head ensures high linearity, high accuracy, and high reliability during measurement.
In order to meet other different measurement tasks, there are various geometric sizes of measuring heads and tips available for selection. Depending on your measurement task, you can use a 2 μ m diamond measuring tip for roughness measurement, a 25 μ m carbon steel measuring tip for contour measurement, or a ruby measuring ball. By using these different probe selections, both short wave and long wave profile components can be recorded and evaluated after a single measurement.
·UD 120/LD 120 measurement process

Simple, fast, and reliable, measuring the workpiece directly in one go yields the following results:
German Mahr MarSurf UD 120 Combination Contour and Roughness Measuring Instrument
·Configuration List
MarSurf UD 120 Measurement Station Shipping List
Order number six million nine hundred and ten thousand and forty-six
Computer, Marsurf XCR 20 software 6268382
Including: Midrange control box and software
Multi language version Win 7 62682xx
19 "TFT LCD display 5460043
Marsurf UD 120 Driver 6720817
Including: measuring arm LD A 14-10-2/60 ° 6852005
Manual control panel MCP 21 7033935
Measurement column ST 500 CNC with HZ adjustment
And marble platform 6710254
Including: Control module 6851376
Probe UD 120/Ld 120
Installed on the measuring column ST 500 6851360
Safety device UD 120/LD 120 7033457
Calibration standard 6820121
X/Y Workbench CT 200 6710530
Seismic pad 6851368
·Technical parameters
Measurement System Optical Measurement System
Travel length 0.1 mm~120 mm
Measuring range of probe: 10 mm~20 mm (using double length measuring arm)
Standard measuring tip Ld a14-10-2/60 °, diamond 2 μ m, 60° (6852005)
Z-resolution of 2 nm
Horizontal measurement axis grating
Point distance in X direction 0.25 μ m~1 μ m
Residual error of 60 nm
Contact force 1 mn~30 mn, adjustable through soft adjustment
Measurement speed of 0.1 mm/s~2.0 mm/s, can be increased in increments of 0.1 mm, used for contour measurement ≤ 0.5 mm/s, used for roughness measurement
Return speed
The maximum positioning speed in the X direction is 4 mm/s
Contact error 0.2 μ m
Distance measurement display error, MPeea ± (1 +2L/150) μm
For example, when measuring a distance of 50 mm, the error is ± 1.66 μ m
Angle measurement error of 1.5 minutes
Radius measurement error MPerr ≤ 10 mm ± 1.5 μ m
10 mm < r ≤ 100 mm ± (3r/20) μm
r > 100 mm ± (7.5 + 9r/40) μm
For example: r 10 mm, The error is ± 1.5 μ m
General
Working environment temperature 15 ° C~30 ° C
Suggested working temperature: 20 ° C ± 2 K
Storage temperature/transportation temperature 0 ° C~40 ° C (in dedicated transport box)
Relative humidity 30%~80%
·Attachment
Simple and accurate arm replacement on MarSurf UD 120/LD 120
For most workpieces, standard measuring heads can be used to measure contours and roughness at various measurement positions. However, not all measurement tasks can be completed with one measuring arm, such as measurements in small holes, measurements in deep grooves, bidirectional measurements, and other measurement tasks. At this point, replacing the measuring arm is also very simple. No tools are needed, the magnetic clamping measuring arm can be easily removed by hand and replaced with another measuring arm in just a few seconds. A spherical locator can ensure precise positioning in micrometers.
As long as the replaced measuring arm has been calibrated, the calibration parameters have been saved, which means that there is no need to recalibrate after replacing the measuring arm!
From the perspective of easy operation and time-saving, this is a substantial advantage of Marsurf UD 120!
German Mahr MarSurf LD 120 Combination Contour and Roughness Measuring Instrument
·MarSurf LD 120 Measurement Station Shipping List
Order number 6910046
Computer with Marsurf XCR 20 software 6268382
Including: Midrange controller and software
Multi language version of Win XP 62682xx
19 "TFT LCD display 5460043
Marsurf UD 120 Driver 6720814
Including: measuring arm
Manual control panel MCP 21 7033935
ST 500 CNC Column 6710254
Equipped with HZ adjustment and marble tablet
Including: Control module 6851376
UD 120/LD 120 probe
Installed on ST 500 measuring column 6851360
UD 120/LD 120 safety protection device 7033457
Calibration standard 6820121
X/Y worktable CT 200 6710530
Seismic pad 6851368
·Technical parameters
Measurement System Optical Measurement System
Travel length 0.1 mm~120 mm
Measuring range of probe: 10 mm~20 mm (using double length measuring arm)
Standard measuring tip Ld a14-10-2, Diamond 2 μ m, 90° (6852002)
Ld a 14-10-500, Ruby measuring ball, 500 μ m (6852003)
Z-resolution of 2 nm
Horizontal measurement axis grating
Point spacing in X direction 0.05 μ m~1 μ m
Residual error of 35 nm
Contact force 0.5 mn~30 mn, adjustable by soft adjustment
Measurement speed of 0.5 mm/s~2.0 mm/s, can be increased in increments of 0.1mm, used for contour measurement ≤ 0.5 mm/s, used for roughness measurement
Return speed
The maximum positioning speed in the X direction is 4 mm/s
Contact error 0.1 μ m
Distance measurement display error, MPeea ± (1+L/100) μ m
For example, when measuring a distance of 50 mm, the error is ± 1.5 μ m
Angle measurement error of 0.5 min
Radius measurement error MPerr ≤ 10 mm ± 1.5 μ m
10 mm < r ≤ 300 mm ± (r/10) μm
r > 300 mm ± (30 + r/5) μm
For example: r 10 mm, The error is ± 1.5 μ m
General
Working environment temperature 15 ° C~30 ° C
Suggested working temperature: 20 ° C ± 2 K
Storage temperature/transportation temperature 0 ° C~40 ° C (in dedicated transport box)
Relative humidity 30%~80%
·MarSurf LD 120 Accessories
MarSurf LD 120 brings the highest accuracy
As shown in the technical parameter values of LD 120 listed earlier, with a residual error value of 35nm, a resolution of 2nm, a stable force measurement of 0.5-30mN, and a minimum measurement point distance of 0.05 μ m, the LD120 probe enables precise measurements on micro surfaces and edge chamfers. But for customers, what is more crucial is whether the ordered equipment can pass the acceptance smoothly, and whether the equipment's capabilities can meet the requirements of the measurement task.
Marsurf LD 120 has been tested and proven by hundreds of customers worldwide. The optimized interaction of mechanical technology, optical technology, control technology, and software technology ensures high measurement accuracy, and you can reliably control dimensional errors to the micrometer level through the measurement of this device.
Application examples