Instructions for using chlorine dioxide test box
Specification Introduction:
Product specification: 0.05-2mg/l
Testing frequency: 30 times/box
Shelf life: 2 years
Storage conditions: Room temperature, cool and dry place
Operating instructions:
Test range: 0.05-0.10-0.25-0.50-0.75-0.00-1.50-2.00mg/L
Testing method:
Rinse the colorimetric tube twice with the water to be tested, and then take a water sample to the mark on the tubeAdd a small glass spoon of chlorine dioxide reagent, shake until the reagent is completely dissolved, let it stand for 3 minutes, and visually compare the color with the standard color chart from top to bottom. The color code that is similar to the color tone of the solution in the tube is the chlorine dioxide content (mg/L) in the water sample to be tested. If observed by single eye color comparison, the effect will be better.
Common interference:
1. PH value. It should be controlled within 5-8.
2. Oxidants: bromine, iodine, aniline, iodoanine, ozone, hydrogen peroxide, chromate, manganese oxide, etc.
3. Reductive agents: Nitrite, etc.
4. If the alkalinity of water exceeds 250mg/l and the acidity exceeds 150mg/l, the measured values will be unstable during the measurement process, and the sample color will disappear or become very light. Adding diluted HCl or NaOH to the sample for neutralization reaction can solve this problem.
Packing list:
4 packs of chlorine dioxide reagents, 1 test tube, 1 dropper, 1 color chart, and 1 glass spoon.
matters needing attention:
This product is non-toxic, but please keep it out of reach of children.
Accidentally touching the skin, rinse off with water.
Other rapid testing kit products of the company:
| model | product name | measuring range | Number of detections | |
| LH2001 | DPD Ozone Determination Kit | 0.05-1mg/l | 50 times/box | |
| LH2002 | DPD residual chlorine determination kit | 0.05-1mg/l | 50 times/box | |
| LH2003 | Export residual chlorine determination kit | 0.1-2mg/l | 50 times/box | |
| Export residual chlorine determination kit | 0.1-2mg/l | 100 times/box | ||
| Export residual chlorine determination kit | 0.1-2mg/l | 300 times/box | ||
| LH2004 | DPD Total Chlorine Determination Kit | 0.05-1mg/l | 50 times/box | |
| LH2006 | Chlorine dioxide determination kit | 0.05-2mg/l | 30 times/box | |
| LH2007 | Ammonia nitrogen determination kit | 0.01-1.0mg/l | 25 times/box | |
| LH2008 | Sulfide determination kit | 0.02-0.8mg/l | 30 times/box | |
| LH2009 | Nitrite determination kit | 0.01-0.5mg/l | 35 times/box | |
| LH2011 | Soft water hardness test kit | 0.4-20mg/l | 50 times/box | |
| LH2012 | Total Hardness Test Kit LR | 4-200mg/l | 50 times/box | |
| LH2013 | Total Hardness Test Kit HR | 12-600mg/l | 50 times/box | |
| LH2014 | Pure Water Nitrite Test Kit | 0.002mg/l (semi quantitative) | 25 times/box | |
| LH2015 | Chloride ion determination kit | 20-400mg/L titration method | 50 times/box | |
| LH2016 | Sulfite determination kit LR | 1-20mg/l | 50 times/box | |
| LH2018 | Alkalinity Test Kit LR | 10-200mg/L, titration | 50 times/box | |
| LH2019 | Alkalinity Test Kit HR | 100-2000mg/L, titration | 50 times/box | |
| LH2020 | Phosphorus determination kit | 0.03-0.5mg/l | 25 times/box | |
| LH2021 | Total iron determination kit | 0.05-1mg/l | 25 times/box | |
| LH2022 | Chromium (VI) determination kit | 0.05-1.0mg/l | 25 times/box | |
| LH2023 | Copper determination kit | 0.2-5mg/l | 25 times/box | |
| LH2024 | Manganese determination kit | 0.1-10mg/l | 25 times/box | |
| LH2025 | Nickel testing box | 0-0.40mg/l | 25 times/box | |
| LH2026 | Arsenic assay kit | 0.004-0.08mg/l | 20 times/box | |
| LH2028 | Aluminum test box | 0-0.20mg/l | 50 times/box | |
| LH2031 | Fluorine reagent kit | 0-1.5mg/l | 50 times/box | |
| LH2032 | Urea reagent kit | 0.5-8mg/l | 50 times/box |