CLJ-3106LHigh flow laser dust particle counter
This instrument is a successful product developed jointly with Nanjing University of Science and Technology and other units in accordance with the requirements of the new version of GMP (2010 version). The overall body of the product is made of 316L stainless steel shell. The laser sensor adopts a full semiconductor laser and a semiconductor photodiode receiver to ensure the stability of the light source and the accuracy of the signal receiver. Greatly reducing the stray light inside the scattering cavity and improving the signal-to-noise ratio of the sensor. Adopting LCD large screen display and PVC film switch control, the display data is clear and the operation is convenient. This instrument can automatically determine various purification levels based on the standards selected by the customer. This product is widely used in medical, pharmaceutical, electronic, chemical, optical, food, cosmetics, biological products, aerospace, and various testing institutions for the detection of dust particles
Main technical parameters:
light source |
Semiconductor laser, with a lifespan of over 10 years>30000 hours |
sample flow |
28.3L/min (1ft3) |
Display mode |
122 × 90mm high-resolution LCD screen display (LCD) with Chinese and English interface selection, real-time display, previous cycle display, real-time concentration display, can display time, date, measurement value, temperature and humidity, room number, sampling point, sampling frequency, status and other parameters, 95% UCL calculation, can directly display particle concentration (particles/cubic meter) |
power supply |
AC power adapter, AC: 100V~245V, 50/60 Hz to DC: 16.8V, 1A; |
working hours |
10 hours |
count mode |
Cumulative value, difference value, concentration value |
testing mode |
Single, repetitive, continuous, computational, remote |
Units |
The unit can be converted into m3 |
Particle size channel |
0.3 、 0.5、 1.0、3.0、 5.0、 10.0μm, Simultaneously count six particle sizes Particle size can be customized according to customer needs |
sampling period |
1~9999 (S) Delay count: 0~99 (S) Self purification time: ≤ 8 (min) |
work environment |
temperature:0~40℃(50~104℉)Relative temperature:20~90%RH, No condensation Atmospheric pressure: 86-106KPa |
Temperature and humidity: |
Choose (1) Temperature: 10-35 ℃± 1 ℃ (2) Humidity: 10-75% RH ± 5% |
UCL settings |
Sampling points (A): Set from 2 to 9 points. Sampling frequency per point: (L) 3-5 times Set measurement position: 0-999 |
UCL report |
Compliant with ISO14644-1, GB50073-2001, GMP (2010) |
Data Storage |
Can store 1000 sets of data(Circular buffer)(including particle size, data, environmental data, year, month, day, time, sampling volume, data location)Data will not be lost after power failure. |
communication interface |
RS232, 9600 baud rate |
Alarm settings |
The instrument is equipped with a level alarm function, which can alarm when the cleanroom exceeds the standards of level 100, 1000, 10000, 100000, 300000, and 1000000. At the same time, it meets the static and dynamic standards of A/B/C/D in the new version of GMP standards, and can switch between four standards for measurement |
zero count |
Complies with JJF-1190-2008 Dust Particle Counter Measurement Calibration Specification and JIS B9921:1997 with less than 1 count per five minutes |
Basis Set Superposition Error |
At 5%, 2000000 particles per cubic foot |
Printing function |
Built-in printer |
weight |
8.6kg |
calibration |
JJF-1190-2008 Dust Particle Counter Metrology Calibration Specification can be traced back to the National Institute of Standards and Technology (NIST) in the United States. Our company has passed the national metrology standard building assessment and can be traced back to the Shanghai Institute of Metrology and Measurement Technology. We can also calibrate ourselves or third-party national metrology institutions |
Standard configuration |
Professional power sampling head, sampling bracket with sampling tube, zero filter, USB to RS 232 module, RS 232 connection cable, computer communication software CD, built-in printer, , |
External dimensions |
350×140×280 |