BINDER FD23 drying oven
1、 Product Introduction:
1. Temperature range: ambient temperature plus 5 ℃ to 300 ℃
2 APT.line ™ Preheating chamber technology
3. Circulating air
4. Adjustable exhaust valve
5. Controller with time function
6. 2 chrome plated sockets
7. Independent adjustable temperature safety device level 2 (DIN 12880) with optical alarm
2、 Technical parameters:
PRODUCT MODEL |
FD023-230V1 | FD023UL-120V1 |
Choose a model | standard | standard |
Item code | 9010-0194 | 9010-0196 |
measure | ||
Internal volume [L] | 20 | 20 |
Net weight of equipment (unloaded) [kg] | 27 | 27 |
Maximum total load [kg] | 25 | 25 |
Maximum load per partition [kg] | 12 | 12 |
The dimensions of the casing do not include additional components and connectors | ||
Net weight width [mm] | 435 | 435 |
Net weight at height [mm] | 495 | 495 |
Depth net weight [mm] | 520 | 520 |
Wall distance from rear [mm] | 160 | 160 |
Wall to side distance [mm] | 100 | 100 |
Internal dimensions | ||
Width [mm] | 222 | 222 |
Height [mm] | 330 | 330 |
Depth [mm] | 300 | 300 |
The number of doors | ||
Outer door | 1 | 1 |
Data related to the environment | ||
Energy consumption at 150 ℃ [Wh/h] | 300 | 300 |
firmware | ||
Number of partitions (hours/maximum) | 2/4 | 2/4 |
Temperature performance data | ||
The temperature range exceeds the indoor temperature by 5 ℃ to [℃] | 300 | 300 |
Temperature deviation is 150 ℃ [± K] | 2.5 | 2.5 |
Temperature fluctuation at 150 ℃ [± K] | 0.3 | 0.3 |
Heating time to 150 ℃ [min] | 25 | 25 |
The recovery time after opening the door in 30 seconds is 150 ℃ [min] | 6 | 6 |
Ventilation data | ||
Air exchange rate at 150 ℃ [x/h] | 64 | 64 |
Electric data | ||
Rated voltage [V] | 230 | 120 |
Power frequency [Hz] | 50/60 | 60 |
Rated power [kW] | 0.8 | 0.8 |
Equipment safety device [A] | 10 | 12.5 |
Phase (rated voltage) | 1~ | 1~ |
1·All technical data is only applicable to standard specifications of unloaded equipment under ambient temperature of 22 ± 3 ℃ and ± 10% power supply voltage fluctuation. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.