BINDER ED260 drying oven
1、 Product Introduction:
1. Temperature range: ambient temperature plus 5 ℃ to 300 ℃
2. Up to 30% more energy-efficient than common devices on the market
3. Utilizing APT.line ™ Technology achieves high temperature accuracy
4. Natural convection
5. Controller with LCD display
6. Exhaust valve electromechanical control
7. 2 chrome plated sockets
8. Integrated independently adjustable temperature safety device level 2 (DIN 12880) with optical alarm
9. Ergonomic handle design
10. USB interface for recording data
2、 Technical parameters:
PRODUCT MODEL | ED260-230V1 | ED260UL-240V1 |
Choose a model | standard | standard |
Item code | 9010-0339 | 9010-0340 |
measure | ||
Internal volume [L] | 255 | 255 |
Net weight of equipment (unloaded) [kg] | 85 | 85 |
Maximum total load [kg] | 270 | 270 |
Maximum load per partition [kg] | 40 | 40 |
The dimensions of the casing do not include additional components and connectors | ||
Net weight width [mm] | 810 | 810 |
Net weight at height [mm] | 965 | 965 |
Depth net weight [mm] | 760 | 760 |
Wall distance from rear [mm] | 160 | 160 |
Wall to side distance [mm] | 100 | 100 |
Internal dimensions | ||
Width [mm] | 610 | 610 |
Height [mm] | 760 | 760 |
Depth [mm] | 550 | 550 |
The number of doors | ||
Outer door | 1 | 1 |
Data related to the environment | ||
Energy consumption at 150 ℃ [Wh/h] | 355 | 355 |
firmware | ||
Number of partitions (hours/maximum) | 2/8 | 2/8 |
Temperature performance data | ||
The temperature range exceeds the indoor temperature by 5 ℃ to [℃] | 300 | 300 |
Temperature deviation is 150 ℃ [± K] | 1.8 | 1.8 |
Temperature fluctuation at 150 ℃ [± K] | 0.8 | 0.8 |
Heating time to 150 ℃ [min] | 50 | 50 |
The recovery time after opening the door in 30 seconds is 150 ℃ [min] | 20 | 20 |
Electric data | ||
Rated voltage [V] | 230 | 240 |
Power frequency [Hz] | 50/60 | 60 |
Rated power [kW] | 2.25 | 2.45 |
Equipment safety device [A] | 12.5 | 16 |
Phase (rated voltage) | 1~ | 1~ |
1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. We reserve the right to make technical changes.