Shanghai Huanao Technology Co., Ltd
Home>Products>BINDER CB60 CO2 incubator
BINDER CB60 CO2 incubator
BINDER CB60 CO2 incubator
Product details

BINDER CB60 CO2 incubator

1、 Product Introduction:

1. Temperature range: ambient temperature plus 7 ℃ to 60 ℃

2. Humidity range: up to 95% relative humidity

3. Optional O2 adjustment range: 0.2-20 Vol. -% O2 or 10-95 Vol. -% O2 (version O, GO)

4. Automatic sterilization through hot air at 180 ℃

5. Humidification system for culture dishes with condensation protection

6. CO2 mixed gas nozzle with Venturi effect

7. CO2 sensor using infrared technology that can be sterilized with hot air

8. Display through LCD color monitor

9. Lockable door handle

10. Sealed inner door made of safety glass (ESG)

11. Seamless deep drawn inner cavity made of stainless steel

12. Two perforated shelves made of stainless steel

13. Stackable devices using stacking adapters

14. Fault diagnosis system using optical and sound alarms

15. Computer interface: Ethernet

16. Internal measurement value recording and USB interface

17. Zero voltage alarm contact

2、 Technical parameters:

PRODUCT MODEL CB060-230V1 CB060-230V-G1 CB060-230V-O1 CB060-230V-GO1 CB060UL-120V1 CB060UL-120V-G1 CB060UL-120V-O1

CB060UL-120V-GO1

Choose a model standard Equipped with a 4-piece detachable inner door With O2Adjust With O2Adjustable and 4-piece detachable inner door standard Equipped with a 4-piece detachable inner door With O2Adjust With O2Adjustable and 4-piece detachable inner door
Item code 9040-0088 9040-0104 9040-0090 9040-0106 9040-0089 9040-0105 9040-0091 9040-0107
measure
Internal volume [L] 53 53 53 53 53 53 53 53
Net weight of equipment (unloaded) [kg] 60 65 63 68 60 65 63 68
Maximum total load [kg] 30 30 30 30 30 30 30 30
Maximum load per partition [kg] 10 10 10 10 10 10 10 10
The dimensions of the casing do not include additional components and connectors
Net weight width [mm] 580 580 580 580 580 580 580 580
Net weight at height [mm] 720 720 720 720 720 720 720 720
Depth net weight [mm] 545 545 545 545 545 545 545 545
Wall to side distance [mm] 50 50 50 50 50 50 50 50
Wall distance from rear [mm] 100 100 100 100 100 100 100 100
Internal dimensions
Width [mm] 400 400 400 400 400 400 400 400
Height [mm] 400 400 400 400 400 400 400 400
Depth [mm] 330 330 330 330 330 330 330 330
The number of doors
Inner door 1 4 1 4 1 4 1 4
Outer door 1 1 1 1 1 1 1 1
Data related to the environment
Energy consumption at 37 ℃ [Wh/h] 80 80 80 80 80 80 80 80
firmware
Number of partitions (hours/maximum) 2/3 2/2 2/3 2/2 2/3 2/2 2/3 2/2
Temperature performance data
The temperature range exceeds the indoor temperature by 7 ℃ to [℃] 60 60 60 60 60 60 60 60
The temperature deviation is 37 ℃ [± K] 0.3 0.3 0.3 0.3 0.3 0.3 0.3 0.3
Temperature fluctuation at 37 ℃ [± K] 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1
The recovery time after opening the door in 30 seconds is 37 ℃ [min] 4 4 4 4 4 4 4 4
Climate performance data
Humidity range [% relative humidity] 90…95 90…95 90…95 90…95 90…95 90…95 90…95 90…95
CO2Data
CO2Range [Vol. -% CO]2] 0…20 0…20 0…20 0…20 0…20 0…20 0…20 0…20
CO2Measurement technology IR IR IR IR IR IR IR IR
CO2The recovery time after opening the door in 30 seconds is 5 vol. -% CO? [min] 5 5 5 5 5 5 5 5
Electric data
Rated voltage [V] 200…240 200…240 200…240 200…240 100…120 100…120 100…120 100…120
Power frequency [Hz] 50/60 50/60 50/60 50/60 50/60 50/60 50/60 50/60
Rated power [kW] 1 1 1 1 1 1 1 1
Equipment safety device [A] 10 10 10 10 16 16 16 16
Phase (rated voltage) 1~ 1~ 1~ 1~ 1~ 1~ 1~ 1~
O2Data
Standard O2Adjustment range: O2Range [Vol. -% O]2] - - 0.2…20 0.2…20 - - 0.2…20 0.2…20
Optional O2Adjustment range: O2Range [Vol. -% O]2] - - 10…95 10…95 - - 10…95 10…95
O2The recovery time after opening the door in 30 seconds is 5 vol. -% O2[min] - - 8 8 - - 8 8

1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. We reserve the right to make technical changes.


Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!